Remote Trip. A negative-going edge applied to terminal J3-30 ( REMOTE TRIP ) will shut down the power supply, reducing the output voltage to near zero. For minimum pulse duration and timing considerations with respect

to REMOTE RESET , see Table A-1. The following paragraph provides a brief circuit description (see schematic diagram and Figure A-8).

A negative going edge at REMOTE TRIP coupled through opto-isolator (U9) causes one-shot U13B to set the

TRIP/RESET latch (U5A) low. This sets terminal J1-13 ( INHIBIT ) low, thus inhibiting the Pulse Width Modulator of the power supply. It also lights the unregulated indicator on the front panel and generates an unregulated signal from the opto-isolator U3.

The low signal generated by the Trip/Reset Latch is also coupled through opto-isolator U2 and appears at J3-17 as

an OVERVOLTAGE status signal. This signal does not affect the state of the power supply's OVP circuit.

Remote Reset. A negative-going edge applied to terminal J3-29 ( REMOTE RESET ) will return the supply to its initial state following a system-initiated shutdown or an OVP shutdown caused by a temporary over voltage

condition. For minimum pulse duration and timing considerations with respect to REMOTE TRIP see Table A-1

under Remote Control. The following paragraphs provide a brief description of this circuit (see schematic diagram and Figure A-8).

A negative-going pulse applied to terminal J3-29 ( REMOTE RESET ) is coupled through opto-isolator U10.

One-Shot U13A then triggers and resets the TRIP/RESET latch output high. This sets terminal J1-13 ( INHIBIT ) high, thus enabling the power supply's Pulse Width Modulator.

The REMOTE RESET signal will also reset the power supply OVP circuit in the event that an overvoltage

condition has shut down the supply. When a REMOTE RESET signal is present, ONE SHOT U13A goes low, this

will produce an OV CLEAR pulse at terminal J1-12. The OV CLEAR pulse will cause the output of A2U2 to go low thus, resetting the OV FLIP FLOP. When this occurs the output of A2U24D goes high and simultaneously

causes the front panel OV LED to turn off and the OV signal (J1-6) to go high. The OVERVOLTAGE signal to U4B also goes high and enables the PWM of the power supply .

Note

By observing the

 

 

 

status indicator or the power supply's output while applying

OVERVOLTAGE

 

a reset pulse to

 

 

, the user can determine the cause of shutdown. If the output

 

REMOTE RESET

 

returns and

 

goes high immediately, this indicates a controller-initiated

 

OVERVOLTAGE

 

shutdown. If the output takes about one second to return, this indicates that the output voltage had

 

exceeded the OVP trip point. If the OVP circuit trips continually, check the load and/or the trip

 

point setting.

Alternate Method of Remote Control. The REMOTE INHIBIT input, J3-31, provides an alternate method of remote shutdown. By maintaining a low logic level at this input, the supply's output will be inhibited until

REMOTE INHIBIT is returned to its initial high state. The following paragraph provides a brief description of this circuit (see schematic diagram and Figure A-8).

A low logic level applied to terminal J3-31 ( REMOTE INHIBIT ) is coupled through opto-isolator U8 and causes U4B to inhibit the power supply's (PWM) Pulse Width Modulator. If jumper W1 is used (see Figure A-8) while a

REMOTE INHIBIT signal is applied, an OVERVOLTAGE signal will appear at terminal J3-17

OVERVOLTAGE thus, indicating the power supply shut down.

103

Page 103
Image 103
Agilent Technologies 6015A, 6010a, 6011A, 6012B service manual Overvoltage, Remote Reset, 103

6015A, 6012B, 6011A, 6010a specifications

Agilent Technologies, a leader in the field of measurement and analysis, offers a suite of instruments within its 6010 and 6011 series, specifically the 6010A, 6011A, 6012B, and 6015A models. These devices are designed to meet the needs of various industries, including healthcare, environmental monitoring, and materials testing.

The Agilent 6010A is a high-performance spectrometer known for its precision and versatility. It utilizes advanced optical technologies to provide exceptional wavelength accuracy and resolution. This model is particularly useful in laboratories where reliable data is critical, offering a wide spectral range and effective noise reduction features. Its user-friendly interface simplifies complex analyses, making it suitable for both seasoned professionals and newcomers.

Following closely, the Agilent 6011A is recognized for its robust capabilities in laboratory environments. This device incorporates advanced signal processing techniques, enabling high-throughput measurements without compromising on quality. The 6011A is ideal for real-time monitoring applications, ensuring that users can make informed decisions based on accurate, timely data. Its comprehensive software suite is designed to enhance data analysis, allowing for seamless integration with existing laboratory workflows.

The 6012B variant enhances the functionality further by introducing additional features tailored for specific applications. With a focus on flexibility, the 6012B supports multiple measurement modes, including direct and differential detection. This model excels in complex measurements, allowing for greater analytical depth and insights. The built-in calibration options ensure consistent performance, making it a reliable choice for various research and development tasks.

Lastly, the Agilent 6015A model stands out with its leading-edge technology, designed for the most demanding applications. It boasts enhanced sensitivity and an improved dynamic range, making it perfect for trace analysis in challenging environmental samples. The 6015A’s advanced reporting tools provide detailed analytics, helping scientists and researchers interpret results efficiently. Its compact design also makes it suitable for laboratory spaces with limited room, without sacrificing performance.

Together, these models showcase Agilent Technologies' commitment to delivering high-quality, innovative solutions that empower users to achieve their analytical goals effectively and efficiently. Whether in a research, clinical, or industrial setting, the 6010A, 6011A, 6012B, and 6015A continue to set standards in precision instrumentation.