10
Table 2-1. Test Equipment Required
TYPE REQUIRED CHARACTERISTICS USE RECOMMENDED MODEL
Oscilloscope Sensitivity: 1mV
Bandwidth: 20MHz & 100MHz
Input: differential, 50 & 10M
P,T Agilent 1740A
Isolation Transformer 100VA 4KVA minimum T
RMS Voltmeter True rms, 10MHz bandwidth
Sensitivity: 1 mV
Accuracy: 5%
P Agilent 3400A
Logic Pulser 4.5 to 5.5Vdc @ 35mA T Agilent 546A
Multimeter Resolution: 100nV
Accuracy: 0.0035%, 6½ digit
P,A,T Agilent 3456A
CC PARD Test
Current Probe
No saturation at:
6010A 20Adc
6011A 100Adc
6012B 51Adc
6015A 51Adc
Bandwidth: 20Hz to 20MHz
P Tektronix P6303
Probe/AM503 Amp/
TM500 Power Module
Electronic Load* Power range: 1000 watts
Open and short switches
P,A Transistor Devices Model
6010A
Voltage range: 200Vdc
Current range: 20Adc
DLP 130-50-2500
DLR-400-15-2500
6011A
Voltage range: 30Vdc
Current range: 120Adc
DLP 50-150-3000
6012B
Voltage range: 65Vdc
Current range: 55Adc
DLP 130-50-2500
6015A
Voltage range: 200Vdc
Current range: 5Adc
DLR-400-15-2500
CC PARD Test
Resistive Load
Value:
6010A
3.5 ohms >1000W
Accuracy: 1%
6011A
0.058 ohms >1000W
Accuracy 1%
6012B
0.4 ohms >1000W
Accuracy: 1%
Rheostat or Resistor Bank
P,A