Installation

When installing the board, perform the following steps:

a.Remove the top and inner cover of the power supply as discussed in Section 3 under Repair and Replacement.

b.Remove the plate next to the barrier strip on the rear panel of the supply by unscrewing the 2 M3 screws.

c.Insert the already prepared 002 board in the slot closest to the right side (looking from the front panel) of the supply.

d.Use the two M3 screws to connect the rear end of the 002 board to the rear panel of the supply.

e.Attach ribbon cables from the A2 Control Board A2J 1 to A7J1 and A2J2 to A7J2.

f.Replace the inner and outer cover of the supply.

g.Remove 550V label from rear of unit.

Connector Assembly Procedure

The following instructions describe assembly of the mating connector provided to interface the user's system with the option connector, J3. Figure A-1 identifies the parts of the mating connector.

Proceed as follows:

Note: It may be desirable to set up a test interface before final assembly of the mating connector to allow checkout of the system. A mating connector with pins accessible for temporary wiring is available from Agilent Technologies, Agilent part number 1251-4464. If the cable assembly presents RFI or ESD problems, a shielded cable assembly accessory Agilent part number 5060-2890 can be ordered.

a.If a multi-wire cable is being used as opposed to individual wires), remove approximately 1 1/2 inches of cable insulation from the end. Be careful not to cut the insulation on the individual wires.

b.Strip 3/16 inch of insulation from the end of each wire to be used.

c.Insert each wire into a contact pin (1) and crimp firmly.

d.Insert each pin into a proper hole in connector-pin house (2) from rear. Pins will lock into housing when fully inserted.

Note

Once the pins are locked into the connector-pin housing, they are extremely difficult to remove.

 

Therefore, be certain pin is in proper hole before inserting fully.

e.Screw a slotted setscrew (3) partially into a square nut (4) and place in position in connector shield assembly

(6).

f.Place strain relief (5) in position in connector shield assembly (6), just under set screw (3). Be certain that strain relief is oriented as shown in Figure A-1.

g.Place connector pin housing (2) in shield assembly (6) and route cable through cable entrance.

h.Fold connector assembly (6) and secure with three screws.

i.Strain relief set screw (3) can now be adjusted from top of connector to clamp firmly on cable.

j.Clip fasteners (7) onto ends of connector pin housing (2).

k.Connector can now be plugged onto option connector J3 and secured with two screws (8) into the threaded stand-offs on either side of J3.

95

Page 95
Image 95
Agilent Technologies 6015A, 6010a, 6011A, 6012B service manual Installation, Connector Assembly Procedure

6015A, 6012B, 6011A, 6010a specifications

Agilent Technologies, a leader in the field of measurement and analysis, offers a suite of instruments within its 6010 and 6011 series, specifically the 6010A, 6011A, 6012B, and 6015A models. These devices are designed to meet the needs of various industries, including healthcare, environmental monitoring, and materials testing.

The Agilent 6010A is a high-performance spectrometer known for its precision and versatility. It utilizes advanced optical technologies to provide exceptional wavelength accuracy and resolution. This model is particularly useful in laboratories where reliable data is critical, offering a wide spectral range and effective noise reduction features. Its user-friendly interface simplifies complex analyses, making it suitable for both seasoned professionals and newcomers.

Following closely, the Agilent 6011A is recognized for its robust capabilities in laboratory environments. This device incorporates advanced signal processing techniques, enabling high-throughput measurements without compromising on quality. The 6011A is ideal for real-time monitoring applications, ensuring that users can make informed decisions based on accurate, timely data. Its comprehensive software suite is designed to enhance data analysis, allowing for seamless integration with existing laboratory workflows.

The 6012B variant enhances the functionality further by introducing additional features tailored for specific applications. With a focus on flexibility, the 6012B supports multiple measurement modes, including direct and differential detection. This model excels in complex measurements, allowing for greater analytical depth and insights. The built-in calibration options ensure consistent performance, making it a reliable choice for various research and development tasks.

Lastly, the Agilent 6015A model stands out with its leading-edge technology, designed for the most demanding applications. It boasts enhanced sensitivity and an improved dynamic range, making it perfect for trace analysis in challenging environmental samples. The 6015A’s advanced reporting tools provide detailed analytics, helping scientists and researchers interpret results efficiently. Its compact design also makes it suitable for laboratory spaces with limited room, without sacrificing performance.

Together, these models showcase Agilent Technologies' commitment to delivering high-quality, innovative solutions that empower users to achieve their analytical goals effectively and efficiently. Whether in a research, clinical, or industrial setting, the 6010A, 6011A, 6012B, and 6015A continue to set standards in precision instrumentation.