Appendix Service Information
Measurement Techniques
167
Service
Information

Measurement Techniques

Setup for Most Tests

Most tests are performed at the front terminals as shown in Figure A-1. Measure
the dc voltage directly at the (+) and (-) terminals on the front panel.
Figure A-1 Performance Verification Test Setup

Current-Monitoring Resistor

To eliminate output current measurement error caused by the voltage drops
in the leads and connections, connect the current monitoring resistor between
the (-) output terminal and the load as a four-terminal device. Connect the
current-monitoring leads inside the load-lead connections directly at the
monitoring points on the resistor element (see RM in Figure A-1).
DVM,
Scope, or
RMS