Appendix Service Information
General Calibration/Adjustment Procedure
183
Service
Information

Front Panel Voltage and Current Calibration

Note Before attempting to calibrate the power supply, you must unsecure the power supply,
and disconnect all loads from the power supply and connect a DVM across the output
terminals. See “Calibration Overview”, starting on page 62 to unsecure.
In the following pr ocedu re , the Agi l e n t E364 0 A mode l is referenc ed to descr ibe
the calibration procedure as an example, so a different calibration value for
each calibration point may be prompted to be adjusted for your specific model.
1 Turn on the calibration mode.
Turn on the calibration mode by holding down (Calibrate) key as you
turn on the power supply and hold down the key until you hear a long beep.
Make sure that the power supply is in ‘‘CV’’ mode. If the power supply is not
in ‘‘CV’’ mode, an error occurs.
Voltage and OVP Calibration
2 Move down a level to the voltage calibration mode.
3 Select the low-end voltage calibration point.
4 Enter the reading you obtained from the DVM by using the knob and
resolution selection keys.

CAL MODE

voltage cal

V LO 0.2

22

2000

V LO 0.4500

Power
Calibrate
View
Calibrate
View
Calibrate
View
Calibrate
View