Appendix Service Information

Performance Test Record for Your Power Supply

179

Service
Information

Performance Test Record for Your Power Supply

CV Performance Test Record

Test Description Models Actual Result Specifications
Upper Limit Lower Limit
CV Programming Accuracy
@ 0 volts (DVM reading)
all +0.0100 V -0.0100 V
CV Readback Accuracy
@ 0 volts
all DVM +0.0050 V DVM -0.0050 V
CV Programming Accuracy
@ Full Scale (DVM reading)
E3640A/42A/44A +20.0200 V +19.9800 V
E3641A/43A/45A +60.0400 V +59.9600 V
CV Readback Accuracy
@ Full Scale
E3640A/42A/44A DVM + 0.0150 V DVM - 0.0150 V
E3641A/43A/45A DVM + 0.0350 V DVM - 0.0350 V
CV Load Effect (Load
Regulation)
E3640A/42A/44A Maximum change: < 5 mV
E3641A/43A/45A Maximum change: < 9 mV
CV Source Effect (Line
Regulation)
E3640A/42A/44A Maximum change: < 5 mV
E3641A/43A/45A Maximum change: < 9 mV
CV PARD (Normal mode) E3640A/42A/44A < 5 mVp-p / 0.5 mVrms
E3641A/43A/45A < 8 mVp-p / 1 mVrms
Load Transient Response
Time
all < 50 msec