Chapter 4 Remote Interface Reference

The SCPI Status Registers

94

SCPI Status System

Binary Weight
20 = 1
21 = 2
22 = 4
23 = 8
24 = 16
25 = 32
26 = 64
27 = 128
28 = 256
29 = 512
210 = 1024
211 = 2048
212 = 4096
213 = 8192
214 = 16384
215 = 32768
+
QUEStionable Status
Event Register Enable Registers
4
"OR"
Not Used
Not Used
Temperature
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
9
+
"OR"
OPC
Not Used
QYE
DDE
EXE
CME
PON
Not Used
5
+
"OR"
Current
Voltage
Overvoltage
Standard Event
Event Register Enable Registers
Operation Complete
Query Error
Device Depenent Error
Execution Error
Command Error
Power On
0
7
2
3
4
STAT:QUES? STAT:QUES:ENAB <value>
STAT:QUES:ENAB?
*ESR? *ESE <value>
*ESE?
Not Used
QUES
MAV
ESB
Not Used
Not Used
Not Used
RQS
5
6
3
4
Serial Poll(SPOLL) *SRE <value>
*STB? *SRE?
Status Byte
Summary Register Enable Register
Output Buffer
0
1