Installing Your Analyzer

STEP 3. Meet Electrical and Environmental Requirements

5.Ensure there are at least six inches of clearance between the sides and back of either the stand-alone analyzer or the system cabinet.

CAUTION The environmental temperature must be 4 °C less than the maximum operating temperature of the analyzer for every 100 watts dissipated in the cabinet. If the total power dissipated in the cabinet is >800 watts, then you must provide forced convection.

6.Set up a static-safe workstation. Electrostatic discharge (ESD) can damage or destroy electronic components.

static-control table mat and earth ground wire: part number 9300-0797

wrist-strap cord: part number 9300-0980

wrist-strap: part number 9300-1367

heel-straps: part number 9300-1308

• floor mat: not available through Agilent Technologies

1-8

Chapter 1

Page 14
Image 14
Agilent Technologies ES, 8753ET quick start Meet Electrical and Environmental Requirements