Quick Start: Learning How to Make Measurements
Learning to Make Reflection Measurements
Step 2. Make a measurement calibration
Follow these instructions to perform an S11 or reflection
1. Select a calibration kit that is appropriate to your device under test. Press Cal
CAL KIT SELECT CAL KIT . Choose the calibration kit that is appropriate to your test device by pressing the appropriate softkey. For example, if your test device uses
2.Press RETURN twice, CALIBRATE MENU S11
REFLECTION
3.Follow the prompts shown on the analyzer display to connect and measure an open, short, and load on PORT 1 or the REFLECTION port.
Any choice of male/female in the calibration process should always be made for the sex that represents the test port. For example, if the test port had a male,
NOTE | To ensure an accurate error correction, you must connect the calibration | ||
| standards to the adapters or cables that you will include in the actual device | ||
| measurement. | ||
|
|
|
|
|
|
|
|
NOTE | If a mistake is made, standards can be measured more than once before | ||
| pressing |
| . Only the last measurement data is used. |
| DONE | ||
|
|
|
|
Figure 2-9 Connections for an S11 or Reflection 1-Port Error-Correction
4.Press DONE 1-PORT CAL after measuring the three standards.
5.Press Save/Recall.
Chapter 2 |