Remote Operation | Remote Operation Command Details |
FREQ?
Description: Gets the current cal factor frequency value from the sensor.
Syntax: FREQ? + LF
Return Value: Current cal factor frequency in GHz
FREQ
Description: Sets the current cal factor frequency value for the sensor.
Syntax: FREQ fghz + LF
Return Value: OK or ERR
Remarks: “fghz” is the cal factor frequency value in GHz. “fghz” must be between 0.05 GHz to 6 GHz. Available resolution is 3 digits after the decimal point.
ZERO
Description: Zeros the power sensor. Syntax: ZERO + LF
Return Value: OK if successful and ERR if zeroing failed.
Remarks: In case of zero failure, the STATUS? command can be used to retrieve more detail about the error. See the STATUS? command for more detail. It takes 19 seconds to zero a sensor, please wait at least this long to get a response from the sensor.
TMP?
Description: Gets the current temperature reading from the sensor.
Syntax: TMP? + LF
Return Value: Current temperature in degrees C.
STATUS?
Description: Get error status byte from the sensor.
Syntax: STATUS? + LF
Return Value: Error status byte
Remarks: Error status byte information:
Status.b0
Status.b1
Status.b2
Status.b3
Status.b4
Status.b5
Status.b6
Status.b7