Remote Operation

Remote Operation Command Details

FREQ?

Description: Gets the current cal factor frequency value from the sensor.

Syntax: FREQ? + LF

Return Value: Current cal factor frequency in GHz

FREQ

Description: Sets the current cal factor frequency value for the sensor.

Syntax: FREQ fghz + LF

Return Value: OK or ERR

Remarks: “fghz” is the cal factor frequency value in GHz. “fghz” must be between 0.05 GHz to 6 GHz. Available resolution is 3 digits after the decimal point.

ZERO

Description: Zeros the power sensor. Syntax: ZERO + LF

Return Value: OK if successful and ERR if zeroing failed.

Remarks: In case of zero failure, the STATUS? command can be used to retrieve more detail about the error. See the STATUS? command for more detail. It takes 19 seconds to zero a sensor, please wait at least this long to get a response from the sensor.

TMP?

Description: Gets the current temperature reading from the sensor.

Syntax: TMP? + LF

Return Value: Current temperature in degrees C.

STATUS?

Description: Get error status byte from the sensor.

Syntax: STATUS? + LF

Return Value: Error status byte

Remarks: Error status byte information:

Status.b0 -> ZERO_TEMP_ERROR (Temperature changed more than allowable limit after zeroing sensor)

Status.b1 -> Not Used

Status.b2 -> ADC_CH2_OR (Temperature over range)

Status.b3 -> ADC_CH3_OR (Detector A over ranged)

Status.b4 -> ZERO_ERROR_DET_A

Status.b5 -> ZERO_ERROR_DET_B

Status.b6 -> TEMP_ERROR (Temperature beyond operating range)

Status.b7 -> Not Used