Sensor Operational Tests

Linearity Test

7.Set up the test for the second 20 dB range as follows:

a.Remove the 10 dB attenuator from in between the reference sensor and splitter and connect the reference sensor directly to the splitter.

b.Remove the MA24106A from the splitter and connect the 10 dB attenuator between the splitter and the MA24106A power sensor (see Figure 5-3).

c.Turn Off the synthesizer’s RF output and perform a low-level Zero of both the Reference sensor and the MA24106A.

d.Turn On the synthesizer’s RF output.

Figure 5-3.Linearity Test Setup

e.Set the synthesizer output level to +10 dBm and then adjust it until the sensor/meter under test reads as close as possible to the value obtained above in step 6e.

8.Record data for the next 20 dB range

a.Read and record the power indicated by the reference meter in the appropriate place in Table 5-4.

b.Lower the output power level of the synthesizer to +5 dBm.

c.Record the reference meter and the MA24106A power sensor readings in the appropriate place in Table 5-4.

MA24106A UG

5-7

Page 47
Image 47