Sensor Operational Tests | Calibration Factor Test |
5-4 Calibration Factor Test
In this test the calibration factors of the MA24106A are compared against another sensor (referred to in this procedure as the “reference sensor”) with known calibration factor uncertainties. This reference sensor should be calibrated by a reputable standards laboratory which has low published measurement uncertainty. To perform the comparison, both sensors are used to measure the output power of a synthesizer with a high quality attenuator, such as the
Test Procedure
1.Set up the equipment as follows (refer to Figure
a.Connect the reference power sensor to the reference power meter using the appropriate cables.
b.Connect the MA24106A USB cable between the personal computer with the Power Meter application installed and the MA24106A power sensor under test.
c.Launch the Power Meter application.
d.Turn the power on to all of the instruments and allow them to warm up for the amount of time specified in their respective manuals.
e.Reset or Preset all of the instruments.
f.Configure the reference meter and sensor to measure a CW signal.
g.Perform a sensor Zero and a 1 mW reference calibration on the reference sensor and meter per the manufacturer’s instructions.
h.Perform a low level Zero of the MA24106A as follows:
With the MA24106A disconnected from the synthesizer, click the Zero Sensor button on the Power Meter application and wait for the Zeroing Sensor message to close.
i.Connect the attenuator to the output of the synthesizer with the appropriate adapter to the output of the attenuator.
j.Set the synthesizer to +6 dBm and 50 MHz.
Figure 5-1. Cal Factor Test Set Up