6.0Maintenance

Fyrite® INSIGHT

 

Fyrite® INSIGHT

If the sensor’s output is too low to be usable, then the message “Bad Calibration Sensor End of Life, Entry Not Saved” will appear. The sensor will now be marked as being BAD in the DIAGNOSTICS screen.

6. Power OFF the regulator of calibration fixture and remove the CO

1

 

cylinder.

 

Fyrite® Insight Components

2

3

6.0Maintenance

6.1Fyrite® INSIGHT Disassembly

The following section describes how to disassemble the Fyrite INSIGHT to perform necessary periodic maintenance. (See Figure 6.1 for diagram.)

Tools Required:

Medium Phillips screwdriver Procedure:

 

5

To O2

J9

(Pump)

 

 

J11

To gas

(Battery)

4

To gas

fitting

To O2

sensor

cap

1. Unplug all thermocouples from bottom of analyzer.

Fig. 6.1

Front View

Side View

2. Remove battery cover and then remove batteries.

 

 

 

 

 

3.Remove sensor caps, disconnect tubing, and then unplug all sensors.

4.Lay analyzer face down on a padded work surface; and then using

a medium Phillips screwdriver, remove the unit’s four rear-case

screws.

5. Lift rear case from analyzer and set aside.

6. Unplug electrical connector J11 from printed circuit board.

1.

O2 Sensor

4.

Sample Pump

2.

CO Sensor

5.

LCD Screen

3.

CO Sensor Base

 

 

Fyrite® Insight End Plate Connectors

Front

7. Lift printed circuit board from analyzer.

6.AC Power Adapter Jack (Power)

7.Sample Gas Thermocouple Connector (T-Stack)

8.Primary Air Thermocouple (T-Air)

9.USB Connector

10.Differential Pressure Hose (Optional)

11.Draft Hose

12.Sample Gas Hose

6 7 8 9

0 11 12

Back

Fig. 6.2

44

Instruction 24-9460

 

Instruction 24-9460

45

Page 25
Image 25
Bacharach manual 6.0Maintenance 6.1Fyrite INSIGHT Disassembly