Fyrite® INSIGHT
Procedure:
NOTE: The
1 |
| 2 | 3 |
4 |
| 4 | Parts: |
| 3 | ||
|
|
| |
|
| 2 | 1. 3/16" I.D. Tubing |
|
| 1 | |
|
| 0 | 2. Tee |
| 5 | ||
| 3. Barbed Hose Connector | ||
| |||
|
| ||
|
| 4. Bellows | |
|
|
5. Manometer
Figure
1.Assemble the pressure sensor calibration equipment as shown in Figure 5.1, but DO NOT connect the analyzer to the calibration equipment at this time.
2.If not already done, power ON the analyzer and display the CALI- BRATION LIST per Section 5.2.
3.Use the ▲▼ buttons to select PRESSURE
and then press ENTER to display the CALI- BRATE PRESSURE screen.
"Measured" is the pressure value currently being detected by the pressure sensor, while
"Applied is a known value of pressure that-4.02will be applied for calibration purposes.
4.With both the
Fyrite® INSIGHT
5.Connect the hose from the manometer to the +ΔP port and apply a negative pressure to this port by adjusting the bellows for a manom- eter reading of
6.Use the ▲▼ buttons to enter an Applied value that exactly equals the manometer reading.
The calibration range is from
7.Wait until the Measured reading stabilizes, and then press ENTER to calibrate the pres- sure sensor's Measured value to that of the Applied value. The message, "Good Calibra- tion" should briefly appear, followed by the CALIBRATION LIST screen.
8.Remove calibration equipment.
5.5 T-Stack Calibration
This procedure first zeroes and then spans the
The use of an electronic thermocouple simulator is the preferred method of producing the desired calibration temperatures. Alternatively, ice and boiling water baths can be used.
Materials Required:
•Thermocouple simulator
-Range: 0 to 600 °F
-Accuracy: ± 0.5 °F
•(Alternatively)
TS-Zero Procedure:
1.Set the thermocouple simulator to room temperature and plug its output into the
Alternatively: Plug the probe's thermocouple into the
38 | Instruction |
| Instruction | 39 |