Bacharach INSIGHT manual T-StackCalibration, TS-ZeroProcedure, Materials Required

Models: INSIGHT

1 31
Download 31 pages 43.36 Kb
Page 22
Image 22
Procedure:

Fyrite® INSIGHT

Procedure:

NOTE: The unit-of-measure for pressure is selected per Section 3.5.3. In the following procedure inwc is selected, but note that any unit- of-measure can be used for calibration purposes.

1

 

2

3

4

 

4

Parts:

 

3

 

 

 

 

 

2

1. 3/16" I.D. Tubing

 

 

1

 

 

0

2. Tee

 

5

-1

 

3. Barbed Hose Connector

 

-2

 

 

 

 

-3

4. Bellows

 

 

-4

5. Manometer

Figure 5-1 Pressure Sensor Calibration Equipment

1.Assemble the pressure sensor calibration equipment as shown in Figure 5.1, but DO NOT connect the analyzer to the calibration equipment at this time.

2.If not already done, power ON the analyzer and display the CALI- BRATION LIST per Section 5.2.

3.Use the ▲▼ buttons to select PRESSURE

and then press ENTER to display the CALI- BRATE PRESSURE screen.

"Measured" is the pressure value currently being detected by the pressure sensor, while

"Applied is a known value of pressure that-4.02will be applied for calibration purposes.

4.With both the -ΔP and +ΔP ports open to the atmosphere, observe that the current Measured pressure reading should be 0.00 ± 0.01 inwc. If necessary, zero the pressure sensor per Section 4.6 then repeat steps 2 through 4.

Fyrite® INSIGHT

5.Connect the hose from the manometer to the +ΔP port and apply a negative pressure to this port by adjusting the bellows for a manom- eter reading of -4.00.

6.Use the ▲▼ buttons to enter an Applied value that exactly equals the manometer reading.

The calibration range is from -6 to -2 inwc (-15 to -5 mb). An attempt to calibrate outside this range will cause the message "Applied Value High" (or Low) to appear at the bottom of the screen.

7.Wait until the Measured reading stabilizes, and then press ENTER to calibrate the pres- sure sensor's Measured value to that of the Applied value. The message, "Good Calibra- tion" should briefly appear, followed by the CALIBRATION LIST screen.

8.Remove calibration equipment.

5.5 T-Stack Calibration

This procedure first zeroes and then spans the stack-temperature channel to known temperature values.

The use of an electronic thermocouple simulator is the preferred method of producing the desired calibration temperatures. Alternatively, ice and boiling water baths can be used.

Materials Required:

Thermocouple simulator (K-type)

-Range: 0 to 600 °F

-Accuracy: ± 0.5 °F

(Alternatively) ice-water, boiling water, thermometer

TS-Zero Procedure:

1.Set the thermocouple simulator to room temperature and plug its output into the T-STACK connector located at the bottom of the analyzer.

Alternatively: Plug the probe's thermocouple into the T-STACK con- nector located at the bottom of the analyzer. DO NOT attach the probe's gas hose to the analyzer's GAS port; otherwise water will be drawn into the analyzer!

38

Instruction 24-9460

 

Instruction 24-9460

39

Page 22
Image 22
Bacharach INSIGHT manual T-StackCalibration, TS-ZeroProcedure, Materials Required