6.2 Water Trap / Filter

Fyrite® INSIGHT

6.2 Water Trap / Filter

6.2.1 Emptying the Water Trap Chamber

The water trap chamber should be emptied after every test, or when the water condensate approaches the tip of the riser tube.

1.Remove water trap chamber per Figure 6.3.

2.Pour out liquid condensate, and then reassemble trap.

6.2.2Replacing the Filter Element

Replace the filter element when it becomes visibly dirty or becomes satu- rated with water.

Material Required: • Filter element, P/N 07-1644

 

• Small flat blade screwdriver

1.

Remove water trap chamber per Figure 6.3.

2.

Pry apart filter chamber using a small flat-blade screw

 

driver. Remove and discard old filter.

3.

Install new filter and reassemble filter chamber, making

 

sure that surfaces “A” and “B” contact each other.

Fyrite® INSIGHT

6.3 O2 Sensor Replacement

Material Required:

• O2 Sensor (P/N 24-0788)

O2CO

Sensor Sensor

Pry tab upward

to remove the battery door

from the unit.

Fyrite® Rear Panel

Fig. 6.4

 

4. Reassemble trap.

Procedure:

1. After battery door is removed, remove

connector tubing from both sensors.

2. Pull O sensor from its socket and

Slot

O2 Cap

 

Pry apart

FILTER

filter chamber

CHAMBER

using a small

 

flat-blade

 

screwdriver

Pull apart

inserted into

using a

notch

twisting

 

motion.

 

FILTER

DO NOT6.2.1 Emptying the Water Trap ChamberELEMENT rock back

and forth!

WATER

TRAP CHAMBER

“B”

When reassem- bling filter chamber, make sure surfaces “A” and “B” contact each other

“A”

2

remove the O2 cap.

3. Properly dispose of the old sensor. En-

gage the nub on the new sensor within

the slot on the cap's side and twist to

secure the cap and sensor together.

4. Install the cap and sensor unit by

• Aligning the ribs on the sides of the

sensor with the corresponding shape

in the base.

• Inserting the pins into the connec-

tors in the base.

Nub

Sensor

Fig. 6.3

5. Reattach tubing.

Fig. 6.5

46

Instruction 24-9460

 

Instruction 24-9460

47