Fyrite® INSIGHT

sured. The differential pressure between the two areas is now displayed on the PRESSURE screen. If the pressure at the +ΔP port is higher than the -ΔP port, the pressure reading will be positive. If it is lower, the reading will be negative.

4.7 Temperature Measurement

The difference in temperature between two areas can be measured by us- ing the analyzer's two temperature channels and the Temperature Screen. By using the T-Air channel as a reference, the temperature applied to the T-Stack channel will be displayed on the Temperature Measurement screen as differential temperature between the two channels.

Perform a differential temperature measurement as follows:

1.Display the MAIN MENU by pressing the MENU (F2) button. If necessary, press ESC until MENU appears above F2.

2.Use the ▲▼ buttons to select TEMPERA-

TURE. Press ENTER to display the Tem- perature Measurement screen.

3.Install thermocouples in both temperature channel connectors.

4.Before taking a measurement, the tempera-

ture channels may need to be zeroed, if not

already displaying zero with both ther-.1 mocouples exposed to the same condition.

Press ZERO (F2) if necessary.

4.8 Saving Test Data

Up to 100 sets ("snap shots") of combustion test, pressure, or temperature data can be saved in memory, which can later be recalled for viewing from the Memory Directory.

1.First display the screen that contains the data to be saved. In the example to the right, all data associated with the Combus- tion Test RUN screen will be saved.

Fyrite® INSIGHT

2.Press the SAVE (F3) button to save the test data in the next available memory location.

NOTE: When memory is full, the next read- ing will not be saved until space is made available by clearing previ- ously saved data. (Section 4.13)

4.9 Ending a Combustion Test

WARNING! Burn Hazard. Do not touch the probe af- ter removing it from the stack. Allow the probe to cool before handling (about 5 minutes).

1.Remove probe from the flue-gas stream.

2.Allow the pump to run until all combustion gases are flushed from the analyzer as indicated by the O2 reading returning to 20.9%.

4.10 Emptying the Water Trap

The Water Trap / Filter Assembly removes stack-gas condensate, and also prevents soot from contaminating the internal components of the analyzer.

IMPORTANT! Use the Water Trap / Filter Assembly in a vertical posi-

tion with the gas-flow arrow pointing up.

Empty the water trap chamber after each combustion test, or stop the test and empty the chamber if the liquid condensate level approaches the tip of the riser tube.

To empty the trap, first pull apart the two halves of the Water Trap using a slight twisting motion; empty the water trap chamber, and then reas- semble the trap.

After each combustion test, also check the Water Trap's filter element. If it looks dirty, replace the filter per Section 6.2.

26

Instruction 24-9460

 

Instruction 24-9460

27

Page 16
Image 16
Bacharach INSIGHT manual Temperature Measurement, Saving Test Data, Ending a Combustion Test, Emptying the Water Trap

INSIGHT specifications

Bacharach INSIGHT is a cutting-edge platform designed for HVAC professionals and building management teams, providing real-time insights and analytics for optimal system performance. Offering a seamless integration of hardware and software, INSIGHT enhances operational efficiency by transforming traditional data management paradigms.

One of the standout features of Bacharach INSIGHT is its user-friendly interface, which enables users to visualize data in an easily digestible format. The platform presents key metrics, historical trends, and system alerts, empowering users to make informed decisions quickly. This accessibility is crucial for technicians and managers who need to identify potential issues before they escalate, ensuring systems run smoothly and effectively.

Another significant characteristic of Bacharach INSIGHT is its advanced data analytics capabilities. The platform leverages machine learning algorithms to analyze performance data, identify patterns, and predict maintenance needs. This proactive approach minimizes downtime and extends the lifespan of HVAC systems, resulting in cost savings for organizations.

INSIGHT also incorporates IoT technologies, allowing for real-time data collection from various connected devices across an HVAC system. This integration ensures that users have access to accurate and timely information, facilitating immediate responses to any irregularities. The use of cloud technology enables data storage and access from anywhere, promoting collaboration among team members and stakeholders.

The platform is designed with scalability in mind, accommodating small businesses with singular HVAC units as well as large enterprises with complex, multi-site systems. This flexibility means that as organizations grow, Bacharach INSIGHT can adapt to their changing needs without requiring extensive modifications.

Furthermore, user customization options allow for tailored dashboards and alerts, ensuring that each team can focus on the most relevant metrics for their operations. This adaptability contributes to improved user experience, enhancing system usability and engagement.

In summary, Bacharach INSIGHT combines intuitive design, powerful analytics, IoT integration, and scalability to redefine HVAC management. By equipping users with actionable insights and fostering proactive maintenance practices, INSIGHT positions organizations to optimize their energy use, reduce costs, and improve overall system reliability. This innovative platform is pioneering a new era for HVAC professionals, facilitating smarter building management with robust, real-time analytics.