Fyrite® INSIGHT

“Measured” is the current temperature reading, while “Applied” is a known temperature that will be applied for calibration purposes.

4. Set thermocouple simulator to 32 °F (0 °C), and then use the

and buttons to enter an Applied value that exactly equals the setting of the simulator.

Alternatively: Submerge probe tip into an ice-water bath with a ther- mometer, wait several minutes, and then use the and but- tons to enter an Applied value that exactly equals the thermometer reading.

The calibration range is from 32 to 41 °F (0 to 5 °C). An attempt to calibrate outside this range will cause the message “Applied Value High” (or Low) to appear at the bottom of the screen.

5.Wait until the Measured reading stabilizes, and then press ENTER to calibrate the TA-Zero Measured value to that of the Applied value; after which the message “Good Calibration” should briefly appear followed by the CALIBRATE TA-SPAN screen.

TA-Span Procedure:

6. Set thermocouple simulator to 212 °F (100 °C), and then use the and buttons to enter an Applied value that exactly equals the set- ting of the simulator.

Alternatively: Submerge probe tip into a container of boiling water with a thermometer, wait several minutes, and then use the and buttons to enter an Applied value that exactly equals the ther- mometer reading.

The calibration range is from 194 to 230 °F (90 to 110 °C). An at- tempt to calibrate outside this range will cause the message “Bad Calibration Wrong CAL Entry” to appear in the following step.

7.Wait until the Measured reading stabilizes, and then press ENTER to calibrate the TA-Span Measured value to that of the Applied value; after which the message “Good Calibration” should briefly appear followed by the CALIBRATION LIST screen being re-dis- played.

Fyrite® INSIGHT

5.7 CO Sensor Calibration

Material required:

Calibration kit, P/N 24-7059

Gas cylinder: 500 ppm CO in air, P/N 24-0492

To improve the accuracy of the CO reading, we suggest that if the analyzer will be primarily used for flue gas testing, then calibrate using 500 ppm CO. If the analyzer, however, will be primarily used for ambient testing, then calibrate using 100 ppm CO.

Procedure:

1.If not already done, turn ON the analyzer and display the CALIBRATION LIST screen per Section 5.2.

2.Use the buttons to highlight CO, and then press ENTER to display the CALI- BRATE CO screen.

“Measured” is the current CO reading,

4

while “Applied” is a known CO level

 

that will be applied for calibration

 

purposes.

 

3.Attach a 500 ppm CO cylinder to the regula- tor of the calibration fixture.

4.Use the and buttons to enter an Applied value that exactly equals the concentration stamped on the CO cylinder.

The calibration range is from 20 to 1,000 ppm. An attempt to calibrate outside this range will cause the message “Applied Value High” (or Low) to appear at the bottom of the screen.

5.Wait until the Measured reading stabilizes and then press ENTER to calibrate the CO Measured value to that of the Applied value. The message “Good Calibration” should briefly appear.

If the sensor’s output is low, but still usable, then the message “Good Calibration WARNING Low Sensor” will appear. The sensor will now be marked as being Low in the DIAGNOS- TICS screen.

42

Instruction 24-9460

 

Instruction 24-9460

43

Page 24
Image 24
Bacharach INSIGHT manual CO Sensor Calibration, TA-Span Procedure

INSIGHT specifications

Bacharach INSIGHT is a cutting-edge platform designed for HVAC professionals and building management teams, providing real-time insights and analytics for optimal system performance. Offering a seamless integration of hardware and software, INSIGHT enhances operational efficiency by transforming traditional data management paradigms.

One of the standout features of Bacharach INSIGHT is its user-friendly interface, which enables users to visualize data in an easily digestible format. The platform presents key metrics, historical trends, and system alerts, empowering users to make informed decisions quickly. This accessibility is crucial for technicians and managers who need to identify potential issues before they escalate, ensuring systems run smoothly and effectively.

Another significant characteristic of Bacharach INSIGHT is its advanced data analytics capabilities. The platform leverages machine learning algorithms to analyze performance data, identify patterns, and predict maintenance needs. This proactive approach minimizes downtime and extends the lifespan of HVAC systems, resulting in cost savings for organizations.

INSIGHT also incorporates IoT technologies, allowing for real-time data collection from various connected devices across an HVAC system. This integration ensures that users have access to accurate and timely information, facilitating immediate responses to any irregularities. The use of cloud technology enables data storage and access from anywhere, promoting collaboration among team members and stakeholders.

The platform is designed with scalability in mind, accommodating small businesses with singular HVAC units as well as large enterprises with complex, multi-site systems. This flexibility means that as organizations grow, Bacharach INSIGHT can adapt to their changing needs without requiring extensive modifications.

Furthermore, user customization options allow for tailored dashboards and alerts, ensuring that each team can focus on the most relevant metrics for their operations. This adaptability contributes to improved user experience, enhancing system usability and engagement.

In summary, Bacharach INSIGHT combines intuitive design, powerful analytics, IoT integration, and scalability to redefine HVAC management. By equipping users with actionable insights and fostering proactive maintenance practices, INSIGHT positions organizations to optimize their energy use, reduce costs, and improve overall system reliability. This innovative platform is pioneering a new era for HVAC professionals, facilitating smarter building management with robust, real-time analytics.