Fyrite® INSIGHT
1.3 Operational Overview
Pressing the PWR button turns the analyzer ON. Note that there is a
To perform a combustion test, choose a fuel code that corresponds to the fuel being burned by the appliance being tested (Section 4.3), and then press the RUN/HOLD button to place the analyzer into its Run Mode. Begin testing by inserting the analyzer’s probe tube into the
Pressing the RUN/HOLD button during a test will freeze all measured and calculated values at their current levels. Pressing the RUN/HOLD button again resumes testing.
A backlight enables an operator to read the display in
An optional power saver function will shut the analyzer OFF after a set time period of inactivity. This feature is disabled if the CO value is greater than 50 ppm or the O2 value is less than 16.0%.
Press the PWR button for at least two seconds to power the analyzer OFF. Note that there is a 5 second delay before the analyzer actually turns OFF, during which time the unit can be kept ON by pressing the RUN/HOLD button. There is a CO purge feature that keeps the analyzer’s pump running until the measured CO level drops below 50 ppm.
Fyrite® INSIGHT
2.0 TECHNICAL CHARACTERISTICS
The Analyzer measures & displays:
Primary/Ambient Air Temperature | |
Oxygen | 0.1 to 20.9% O2 |
Stack Temperature | |
Carbon Monoxide | 0 to 2000 ppm CO |
Differential Pressure/Draft | ± 100 inwc |
The Analyzer computes & displays: |
|
Combustion Efficiency | 0.1 to 100% |
Carbon Dioxide | 0.1 to a fuel dependent |
| maximum value in % |
CO referenced to %O2 | 0 to 9,999 ppm |
Excess Air | 0 to 250% |
Fuel Selection |
|
• Natural Gas | • Propane |
• Oil #2 | • Kerosene |
• Oil #4 | • Wood |
• Oil #6 | • Coal |
Normal Operating Conditions: |
|
Temperature |
|
• Analyzer | 32 to 104 °F (0 to 40 °C) |
• Probe | 1,000 °F max (538 °C) at |
| 5" insertion |
Humidity | 15 to 90% RH, |
Dimensions | Weight |
7.9 in (20.1 cm) Height | 1lb. (0.45 kg) |
3.6 in (9.1 cm) Width |
|
2.3 in (5.8 cm) Depth |
|
4 | Instruction |
| Instruction | 5 |