Fyrite® INSIGHT
2.Allow the analyzer to cycle through its 60 second
•Oxygen sensor spanned to 20.9%
•CO sensor zeroed
•Pressure sensor zeroed
Any errors detected during
3.Display the MAIN MENU by pressing the MENU (F2) button. If necessary, press ESC until MENU appears above F2.
4.Use the ▲▼ buttons to select CALIBRA-
TION. Press ENTER to display the Calibra- tion Password screen.
5.Before calibration can begin a
6.Press the ENTER button to accept the password. If the correct password was entered the CALIBRATION Menu will be displayed.
7.Use the ▲▼ buttons to select the desired sensor channel to be calibrated, and then perform the calibration procedure for that sensor as described in the following sections.
Fyrite® INSIGHT
5.3B-Smart Sensor Replacement & Calibration
To replace a
1.Enter the CALIBRATION MENU per Sec- tion 5.2.
2.Use the ▲▼ buttons to select
3.Use the ▲▼ buttons to enter the 10 digit al- phanumeric code supplied with the
NOTE: If the correct code was entered, the analyzer accepts it and returns to the CALIBRATION MENU. If an incorrect code was entered, the screen will display "Invalid Code". Check to make sure the correct code has been entered. If problem persists, contact your nearest Bacharach Service Provider.
Bacharach also offers a convenient Exchange Program that allows the cus- tomer to regularly send in old sensors and, as scheduled, receive new re- placements that are already calibrated and that include a code that can be entered into the analyzer for a quick convenient setup. Contact Bacharach customer service for more details about this program.
5.4 Pressure Sensor Calibration
This procedure calibrates the pressure sensor to a known pressure value.
Materials Required:
•Bellows
•Manometer
-Range: ± 8 in. of water column (± 20 mb)
-Accuracy: ± 0.01 in. of water column (± 0.025 mb)
36 | Instruction |
| Instruction | 37 |