Fyrite® INSIGHT
6.4 CO Sensor Replacement
Material Required:
• CO Sensor (P/N 24-7265) or B-Smart sensor (P/N 24-1467)
O2CO
Sensor Sensor
Pry tab upward
to remove the battery door
from the unit.
Fyrite® Rear Panel | Fig. 6.6 |
|
Fyrite® INSIGHT
6.5 Thermocouple Replacement
Using the appropriate thermocouple replacement kit listed below, replace the probe’s thermocouple as follows:
Each kit contains a thermocouple assembly, two
Tools Required:
•Small flat blade screwdriver
•Wire cutter
•Wire stripper
•Slip joint pliers
Procedure:
1.Gain access to the thermocouple connections by first removing three screws from probe handle, and then separating the two handle pieces.
2.Cut wires attached to old crimp connectors, leaving behind as much of the probe’s
Probe Handle
Screws
Procedure:
1.After battery door is removed, remove the connector tubing from the CO sensor.
2.Remove CO cap by twisting counter clockwise. Gently pull CO sensor out of its socket.
3.Properly dispose of the old CO sen- sor.
CO Cap
Sensor
Printed Circuit Board
Fig. 6.8
4.Plug the new CO sensor into its socket.
5.Install the CO cap by aligning the arrow on the cap with the "OPEN" marker, then twisting to the "CLOSED" marker. Reattach tub- ing.
OPEN
CLOSED
Fig. 6.7
3.Pull old thermocouple from probe body and discard.
4.The new thermocouple has been coiled for shipping purposes. Straighten the thermocouple using your thumb and index finger.
5.If not already done, install supplied
6.Insert thermocouple into probe body until it “bottoms out.”