Fyrite® INSIGHT

Low Sensor- O2 or CO sensor outputs were low but still usable. Sensor(s) may need to be replaced in the near future. Message will indi- cate which sensor(s) were in warning.

Warmup Sensor Error -

CO sensor was not zeroed at warmup because of high output. Run instrument on fresh air then restart instrument to re-zero sensor. If message persists, CO sensor may need to be replaced.

Stack or Air temperature channel is measuring temperature outside the range of -4 to 212 ˚F at startup. Make sure that the Stack and Air thermocouples are sampling ambient room air within the tem- perature range at startup.

Pressure sensor is measuring pressure outside the range of ± 3 inwc at startup. Ensure that the analyzer is sampling atmospheric pres- sure and restart the instrument.

The analyzer was turned on with the probe sampling flue gas. Move the probe to fresh air and restart the instrument.

Message will indicate which channel is in error.

Low Battery - Battery voltage is low. Replace the batteries.

XXXX- Occurs in the number fields of sensors that have achieved over- range condition.

****- Occurs in the number fields of sensors and the calculated values that depend on the sensors that were in error coming out or warmup.

-- - - Occurs in the number fields of calculated values when Oxygen is above 16.0%.

6.9 Diagnostics Screen

The DIAGNOSTICS menu provides information regarding the operation of the analyzer. Information includes the following:

Time Meters - Displays the run time of the analyzer and sample pump in hours of operation.

Fyrite® INSIGHT

Main Diagnostics - Lists the current status of the Stack and Air Ther- mocouple channels, Reference Temperature channel, Pressure channel, CO channel, O2 channel and Battery.

O2 Sensor Life - Displays the approximate remaining life of the Oxygen sensor.

Fresh Air Diagnostics - Fresh air diagnostics will cycle the instrument through the 60 second warm-up to check on the status of the sensors. The probe must be in fresh air to perform a valid check.

Access the DIAGNOSTIC menu as follows:

1.Display the MAIN MENU by pressing the MENU (F2) button. If neces- sary, press ESC until MENU appears above F2.

2.Use the ▲▼ buttons to highlight DIAGNOSTICS and then press ENT to display the DIAGNOSTIC MENU.

3.Use the ▲▼ buttons to highlight the desired diagnostic topic and then press ENT to display the information under that topic.

4.Press the ESC key to exit back to the DIAGNOSTIC MENU or the MENU (F2) key to exit back to the MAIN MENU.

6.10 Status Screen

The Status Screen provides a quick reference to key items when trouble- shooting:

Version

ADC Version

Boot Version

Model

SN

Access the Status menu as follows:

1.Display the MAIN MENU by pressing the MENU (F2) button. If neces- sary, press ESC until MENU appears above F2.

2.Use the ▲▼ buttons to highlight STATUS and then press ENT to dis- play the DEVICE STATUS screen.

3.Press the ESC key to exit back to the MAIN MENU.

52

Instruction 24-9460

 

Instruction 24-9460

53

Page 29
Image 29
Bacharach INSIGHT manual Diagnostics Screen, Status Screen

INSIGHT specifications

Bacharach INSIGHT is a cutting-edge platform designed for HVAC professionals and building management teams, providing real-time insights and analytics for optimal system performance. Offering a seamless integration of hardware and software, INSIGHT enhances operational efficiency by transforming traditional data management paradigms.

One of the standout features of Bacharach INSIGHT is its user-friendly interface, which enables users to visualize data in an easily digestible format. The platform presents key metrics, historical trends, and system alerts, empowering users to make informed decisions quickly. This accessibility is crucial for technicians and managers who need to identify potential issues before they escalate, ensuring systems run smoothly and effectively.

Another significant characteristic of Bacharach INSIGHT is its advanced data analytics capabilities. The platform leverages machine learning algorithms to analyze performance data, identify patterns, and predict maintenance needs. This proactive approach minimizes downtime and extends the lifespan of HVAC systems, resulting in cost savings for organizations.

INSIGHT also incorporates IoT technologies, allowing for real-time data collection from various connected devices across an HVAC system. This integration ensures that users have access to accurate and timely information, facilitating immediate responses to any irregularities. The use of cloud technology enables data storage and access from anywhere, promoting collaboration among team members and stakeholders.

The platform is designed with scalability in mind, accommodating small businesses with singular HVAC units as well as large enterprises with complex, multi-site systems. This flexibility means that as organizations grow, Bacharach INSIGHT can adapt to their changing needs without requiring extensive modifications.

Furthermore, user customization options allow for tailored dashboards and alerts, ensuring that each team can focus on the most relevant metrics for their operations. This adaptability contributes to improved user experience, enhancing system usability and engagement.

In summary, Bacharach INSIGHT combines intuitive design, powerful analytics, IoT integration, and scalability to redefine HVAC management. By equipping users with actionable insights and fostering proactive maintenance practices, INSIGHT positions organizations to optimize their energy use, reduce costs, and improve overall system reliability. This innovative platform is pioneering a new era for HVAC professionals, facilitating smarter building management with robust, real-time analytics.