Display function selection

TO SELECT A DISPLAY FUNCTION (DF):

1Open the main menu.

2Open the Display Function menu.

3Select Display Function.

4Select the desired DF.

5Return to the Display Function menu. You can select the following DFs:

DICOM

Select a DICOM display function for most

 

medical viewing applications.

 

The DICOM function results in more visible

 

grayscales in the images.

 

 

Dynamic Gamma 2.2

These are gamma functions that are shifted

Dynamic Gamma 1.8

to take into account the non-zero

 

luminance of an LCD panel when driven

 

with a “black” signal.

 

They are especially useful in CT applications

 

to improve the perception of low

 

Hounsfield values.

 

 

Native

If you select Native, the native panel

 

behavior will not be corrected.

 

 

Test

This is identical to the Native DF.

 

 

Gamma 2.2

Select one of these display functions in

Gamma 1.8

case the display is to replace a CRT display

 

with a gamma of 1.8 or 2.2 respectively.

 

 

User

This is a user-definable DF.

 

 

Coronis Fusion 4MP/6MP 43

Page 43
Image 43