Measured ALC value

The value shown here is taken into account

 

in the calculation of the DICOM DF if

 

Continuous DICOM ALC is switched off and

 

DICOM Offset is set to “Measured ALC”.

 

When you select this line and confirm the

 

selection, the actual averaged ambient

 

light value is stored here, overwriting the

 

former value.

 

However, if the averaged ambient light

 

value is higher than the maximum ambient

 

light value that corresponds to the selected

 

reading room, a warning is displayed

 

prompting you to dim the light in the

 

room, and the value will be limited to the

 

maximum ambient light value.

 

If you dim the light in the room you can

 

execute the Measure ALC Value function

 

again to enter the new, lower value.

 

 

CALIBRATION INFORMATION SUBMENU

 

 

Preset ambient value

Shows the ambient light correction value

/ Measured ALC

taken into account to calculate the DICOM

ambient value /

DF, expressed in lux.

Average ambient

 

value

 

 

 

Bright luminance

Shows the bright luminance value taken

 

into account to calculate the DICOM DF.

 

 

Dark luminance

Shows the dark luminance value taken into

 

account to calculate the DICOM DF.

 

 

Ambient correction

Shows the ambient light correction value

 

taken into account to calculate the DICOM

 

DF, expressed in cd/m².

 

 

Coronis Fusion 4MP/6MP 35

Page 35
Image 35
Barco 4MP/6MP manual Calibration Information Submenu