Operation Notes and Details

WARNING

Discontinue lift use immediately if any lift component does not operate prop- erly. Failure to do so may result in serious bodily injury and/or property damage.

As the UP switch is pressed and the platform raises from ground level, the inner roll stop automati- cally unfolds (rotates) to the hori- zontal (bridging) position when it UHDFKHVYHKLFOHÁRRUOHYHO7KH roll stop must overlap the lift base plate a minimum 1/2".

When the FOLD switch is pressed

DQGWKHSODWIRUPIROGVIURPÁRRU level to the stow (vertical) position, the roll stop/bridge plate automati- cally travels inboard to the stowed position.

Interlock Features: The inner roll stop features a locking mecha- nism that prohibits the platform from lowering if the lock is not engaged. The inner roll stop also senses weight to prohibit lift op- eration. The lift will not function if the inner roll stop is occupied.

Discontinue lift operation im- mediately if the inner roll stop does not operate properly.

Outer Barrier: This spring-load- ed roll stop provides a ramp for wheelchair loading and unload-

ing at ground level (see photos on following page). When the platform lowers fully to ground level, the roll stop activation feet automatically unfold (rotate) the barrier to the ramp position (fully unfolded). The two activation feet (one at each end of the bar- rier) ensure barrier operation on an uneven surface.

Although the outer barrier is lift-powered, the activation of the barrier is controlled by the lift operator (attendant). Press- ing the DOWN switch unfolds the barrier. The outer barrier is spring-loaded to automati- cally fold (rotate) to the vertical position when the UP switch is pressed. As the activation feet lift off the ground, the torsion springs rotate the barrier to the

Page 19

Page 21
Image 21
Braun NVL SERIES manual When the Fold switch is pressed