Compur Statox 4120

DIP switch S1 (see pic. 8) programs gas and measuring range. If you change your Statox to another gas or measuring range, a hardware modification may be required and the electronics must be re - adjusted. This must be done by authorised personnel.

Table 1: Switch position for H2S

Measuring range:

0-30 ppm

0-50 ppm

0-100 ppm

S1/1

ON

OFF

OFF

S1/2

ON

ON

ON

S1/3

ON

ON

ON

S1/4

ON

ON

OFF

S1/5

ON

ON

OFF

S1/6

ON

OFF

ON

S1/7

ON

ON

OFF

S1/8

ON

ON

ON

E-PROM Index sensor head

12, 13, 14

12, 13, 14

12, 13, 14

E-PROM Index control module

≥ 02

≥ 02

≥ 02

Table 2: Switch position for HCN

Measuring range:

0-15 ppm

0-30 ppm

0-50 ppm

0-100 ppm

S1/1

ON

ON

OFF

OFF

S1/2

ON

ON

ON

ON

S1/3

ON

ON

ON

ON

S1/4

OFF

OFF

OFF

ON

S1/5

OFF

ON

ON

ON

S1/6

OFF

ON

OFF

ON

S1/7

OFF

ON

ON

ON

S1/8

ON

ON

ON

ON

E-PROM Index sensor head

13

12, 13, 14

12, 13, 14

15

E-PROM Index control module

≥ 02

≥ 02

≥ 02

≥ 02

Table 3: Switch position for COCl2

 

Measuring range:

 

 

0-0,1 ppm

 

 

0-0,3 ppm

 

 

0-0,3 ppm

 

 

0-0,5 ppm

 

 

0-1,5 ppm

 

 

0-15 ppm

 

 

0-100 ppm

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

S1/1

 

 

OFF

 

 

ON

 

 

ON

 

 

OFF

 

 

OFF

 

 

OFF

 

 

OFF

 

 

S1/2

 

 

ON

 

 

ON

 

 

ON

 

 

ON

 

 

ON

 

 

ON

 

 

ON

 

 

S1/3

 

 

ON

 

 

ON

 

 

ON

 

 

ON

 

 

ON

 

 

ON

 

 

ON

 

 

S1/4

 

 

OFF

 

 

ON

 

 

ON

 

 

ON

 

 

OFF

 

 

ON

 

 

ON

 

 

S1/5

 

 

ON

 

 

OFF

 

 

ON

 

 

OFF

 

 

ON

 

 

OFF

 

 

OFF

 

 

S1/6

 

 

ON

 

 

ON

 

 

ON

 

 

OFF

 

 

OFF

 

 

OFF

 

 

ON

 

 

S1/7

 

 

ON

 

 

ON

 

 

ON

 

 

ON

 

 

OFF

 

 

OFF

 

 

ON

 

 

S1/8

 

 

ON

 

 

ON

 

 

ON

 

 

ON

 

 

ON

 

 

ON

 

 

ON

 

 

E-PROM Index

 

20

 

12, 13, 14

 

18

 

12, 13, 14

 

12, 13, 14

 

12, 13, 14

 

17

 

 

sensor head

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

E-PROM Index

 

04

 

 

02

 

 

02

 

 

02

 

 

02

 

 

02

 

 

02

 

 

control module

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

16

Page 16
Image 16
Compur 4120 operation manual Switch position for H2S, Switch position for HCN, Switch position for COCl2