Compur Statox 4120

DIP switch S1 (see pic. 8) programs gas and measuring range. If you change your Statox to another gas or measuring range, a hardware modification may be required and the electronics must be re - adjusted. This must be done by authorised personnel.

Table 1: Switch position for H2S

Measuring range:

0-30 ppm

0-50 ppm

0-100 ppm

S1/1

ON

OFF

OFF

S1/2

ON

ON

ON

S1/3

ON

ON

ON

S1/4

ON

ON

OFF

S1/5

ON

ON

OFF

S1/6

ON

OFF

ON

S1/7

ON

ON

OFF

S1/8

ON

ON

ON

E-PROM Index sensor head

12, 13, 14

12, 13, 14

12, 13, 14

E-PROM Index control module

≥ 02

≥ 02

≥ 02

Table 2: Switch position for HCN

Measuring range:

0-15 ppm

0-30 ppm

0-50 ppm

0-100 ppm

S1/1

ON

ON

OFF

OFF

S1/2

ON

ON

ON

ON

S1/3

ON

ON

ON

ON

S1/4

OFF

OFF

OFF

ON

S1/5

OFF

ON

ON

ON

S1/6

OFF

ON

OFF

ON

S1/7

OFF

ON

ON

ON

S1/8

ON

ON

ON

ON

E-PROM Index sensor head

13

12, 13, 14

12, 13, 14

15

E-PROM Index control module

≥ 02

≥ 02

≥ 02

≥ 02

Table 3: Switch position for COCl2

 

Measuring range:

 

 

0-0,1 ppm

 

 

0-0,3 ppm

 

 

0-0,3 ppm

 

 

0-0,5 ppm

 

 

0-1,5 ppm

 

 

0-15 ppm

 

 

0-100 ppm

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

S1/1

 

 

OFF

 

 

ON

 

 

ON

 

 

OFF

 

 

OFF

 

 

OFF

 

 

OFF

 

 

S1/2

 

 

ON

 

 

ON

 

 

ON

 

 

ON

 

 

ON

 

 

ON

 

 

ON

 

 

S1/3

 

 

ON

 

 

ON

 

 

ON

 

 

ON

 

 

ON

 

 

ON

 

 

ON

 

 

S1/4

 

 

OFF

 

 

ON

 

 

ON

 

 

ON

 

 

OFF

 

 

ON

 

 

ON

 

 

S1/5

 

 

ON

 

 

OFF

 

 

ON

 

 

OFF

 

 

ON

 

 

OFF

 

 

OFF

 

 

S1/6

 

 

ON

 

 

ON

 

 

ON

 

 

OFF

 

 

OFF

 

 

OFF

 

 

ON

 

 

S1/7

 

 

ON

 

 

ON

 

 

ON

 

 

ON

 

 

OFF

 

 

OFF

 

 

ON

 

 

S1/8

 

 

ON

 

 

ON

 

 

ON

 

 

ON

 

 

ON

 

 

ON

 

 

ON

 

 

E-PROM Index

 

20

 

12, 13, 14

 

18

 

12, 13, 14

 

12, 13, 14

 

12, 13, 14

 

17

 

 

sensor head

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

E-PROM Index

 

04

 

 

≥ 02

 

 

≥ 02

 

 

≥ 02

 

 

≥ 02

 

 

≥ 02

 

 

≥ 02

 

 

control module

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

16

Page 16
Image 16
Compur 4120 operation manual Switch position for H2S, Switch position for HCN, Switch position for COCl2