5.1 Using the diagnostic box

Switch the diagnostic box on. Then connect it to the opto- coupler receptacle on sensor head bottom. Keep the LED`s and the phototransistors of the opto- coupler clean for proper data transmission.

5.1.1 Operation and error codes of the diagnostic box

Note: If a combination of push button operations is requested ( f. i. ST + ENT), the buttons must be operated in a short time distance.

Example: ST + ENT initiates a self test. The display says „WAIT-TEST RUN“ until the self test is finished.

Diagnostic box queries:

 

 

Button

Display

 

 

 

M

Measured value in ppm

 

 

 

M + ENT

WAIT; after a while reading of actual measured value in ppm

 

 

 

RA

Sensor head measuring range

 

 

 

RA + ENT

Gas

 

 

 

ST

Status passed (System OK) or error message (see 5.1.2)

 

 

 

ST + ENT

(initiates self test), TEST Run

 

 

 

FKT

Calibration factor

 

 

 

ZERO + ENT *

ZERO RUN (start zero adjustment)

 

 

 

* only available with systems with zero adjustment

Diagnostic box error messages:

 

 

 

 

 

 

ERR INPU

 

 

Faulty or too slow input. Repeat.

 

 

 

 

 

 

ERR KONZ

 

 

Calibration gas concentration out of range.

 

 

 

 

 

 

ERR FACT

 

 

Calibration factor is out of range.

 

 

 

 

 

 

DS.W.POS

 

 

Measuring range or gas not correctly programmed.

 

 

 

 

 

 

WARN TIM

 

 

Time error during data transmission.

 

 

 

 

 

 

WARN DTE

 

 

Faulty data transmission.

 

 

 

 

 

 

WARN CHE

 

 

No self test due to too low temperature.

 

 

 

 

 

 

TEMP.OU

 

 

Temperature out of the admissible range.

 

 

 

 

 

 

TEMP.IN

 

 

Temperature is back in the admissible range.

 

 

 

 

 

 

 

CH.BATT

 

 

Charge calibration box battery.

 

(audible alarm)

 

 

 

 

 

 

 

 

NO ZERO

 

 

Zero adjustment not possible. Wrong EPROM in sensor

 

 

 

head and control module.

 

 

 

 

 

 

PUMP RUN

 

 

Sensor is flushed with air because the gas concentration is

 

 

 

too high.

 

 

 

 

 

 

CAL END

 

 

Calibration routine has been cancelled with CE button.

 

 

 

 

 

 

NO CALIB

 

 

Calibration not possible within the time limit.

 

 

 

 

 

 

 