7. The maintenace request option

Purpose:

This option can differentiate between fatal errors in the sensor head and maintenance requirements. It requires a Statox 4120 control module equipped with EPROM Index ≥ 03. It also sets the analog output to 4 mA during the self test, if required.

Operation modes:

The Statox 4120 can operate in 9 different modes. The relevant mode can be seen by the control module LED`s and relays.

Operation mode Statox

Analog output

SR-LED

SF-LED

SF-Relay

MR-

PCS-*

 

 

Control

Control

 

Relay

Relay

 

 

module

module

 

 

 

O Self test

0mA / 4mA / 0V

„on“

„on“

active

passive

active

 

 

 

 

 

 

 

O Diagnostic box connected:

0mA / 4mA / 0V

flashing

„off“

active

passive

active

 

 

 

 

 

 

 

O Alarm threshold adjustment:

0mA / 4mA / 0V

„on“

„off“

active

passive

active

 

 

 

 

 

 

 

O Start-mode

4mA / 0V

flashing

„off“

active

passive

passive

 

 

 

 

 

 

 

O Detection mode:

Analog signal

„on“

„off“

active

passive

passive

 

 

 

 

 

 

 

O Maintenance request:

 

 

 

 

 

 

B1 Battery Low

 

 

 

 

 

 

B2 Battery low under load

 

 

 

 

 

 

C1 Purge before test

 

 

 

 

 

 

C2 Purge between test peaks

 

 

 

 

 

 

C3 Sensitivity

Analog signal

flashing

„off“

active

active

passive

(programmable by S3)

 

 

 

 

 

 

C4 Purge

 

 

 

 

 

 

PU Pump current too high

 

 

 

 

 

 

T2 Temperature out of specs

 

 

 

 

 

 

 

 

 

 

 

 

 

O System failure:

 

 

 

 

 

 

A1 Amplifier

4mA / 0V

„off“

„on“

open

passive

passive

C3 Sensitivity

 

 

 

 

 

 

(programmable by S3)

 

 

 

 

 

 

 

 

 

 

 

 

 

O Data transmission error

4mA/0V

„off“

„on“

open

passive

passive

 

 

 

 

 

 

 

O Power failure

0mA/0V

„off“

„off“

open

passive

passive

 

 

 

 

 

 

 

*(PCS=Process control system)

25

Page 25
Image 25
Compur 4120 operation manual Maintenace request option