5.1.2 Status messages of the sensor head Button ST scans the system status:

Code

Status and action required

 

 

Passed

System OK.

Error A1

Amplifier defective: Repeat adjustment or replace pcb.

Error B1

Check battery pack connection. Battery pack discharged or defective, replace.

 

If battery pack discharges frequently, check charging circuit.

Error B2

Battery pack voltage drops under load.

Error PU

Pump energy consumption too high. Replace.

Error T2

Temperature is out of compensation range. Measured values are corrected

 

with the end of range - value.

Error C1

Zero not found at start of self test (or sensor signal is too high for zero

 

adjustment).

 

- check filter cartridge: Fresh? Tight?

 

- Filter cartridge gas intake at the bottom of the sensor head dirty

 

- Filter cartridge exhausted

 

- Dust filter dirty

 

- Sensor too slow

Error C2

Zero not found at end of self test:

 

Sensor too slow, replace.

Error C3

Test peak not achieved:

 

- Sensor too slow

 

- Sensor lost sensitivity

 

- Gas generator expired

 

- Sensor too slow

 

- Pump defective

 

- Filter cartridge polluted

Error C4

Zero not found at end of self test:

 

- Sensor too slow, replace.

 

- Filter cartridge polluted

 

- Pump defective

Caution:

Disconnect the diagnostic box first and then switch it off. Don not forget to protect the opto - coupler with the plug. To preserve battery power, the diagnostic box display goes off after 2 minutes. Operate any key to switch it on again.