Compur Statox 4120

10. Technical data General:

Gas

Measuring range

Cl2

0-1.5 ppm, 0-3 ppm, 0-10 ppm

CO

0-150 ppm

COCl2

0-0.1 ppm, 0-0.3 ppm, 0-0.5 ppm, 0-1.5 ppm, 0-15 ppm, 0-100 ppm

HCN

0-15 ppm, 0-30 ppm, 0-50 ppm, 0-100 ppm

HCl

0-100 ppm

H2S

0-30 ppm, 0-50 ppm, 0-100 ppm

ClO2

0-0.5 ppm

NO2

0-15 ppm

SO2

0-5 ppm

Accuracy at TLV:

+/- 10 %

Alarms:

2 adjustable alarm thresholds, system alarm

RFI:

CE conform

Manufacturer:

Compur Monitors, Munich

Sensor head:

Type:

5330 xxx (xxx: measuring range / gas)

Weight:

1.9 kg

Dimensions (mm, HxBxT):

180x300x180

Operating temperature:

-20 to +40 oC

Storage temperature:

-30 to +50oC

Humidity:

20-95 % r.H. (not condensing)

Pressure:

800-1200 hPa

Protection class:

IP53

Approvals (div.countries):

Cl2, COCl2, H2S

Explosion protection:

EEx ib IIC T6

Operating environment :

II 2 G

Operating voltage Ui:

max. 22 VDC

Operating current Ii:

max. 50 mADC

Internal capacity Ci:

55 nF

Internal inductivity Li:

0 mH

30

Page 30
Image 30
Compur 4120 operation manual Technical data General Gas Measuring range, Sensor head