Compur Statox 4120

Table 4: Switch position for Cl2

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Measuring range:

0-1,5 ppm

 

0-3 ppm

 

0-10 ppm

 

S1/1

ON

 

ON

 

OFF

 

S1/2

ON

 

ON

 

ON

 

S1/3

ON

 

ON

 

ON

 

S1/4

ON

 

OFF

 

ON

 

S1/5

ON

 

ON

 

OFF

 

S1/6

ON

 

ON

 

ON

 

S1/7

OFF

 

OFF

 

OFF

 

S1/8

ON

 

ON

 

ON

 

E-PROM Index sensor head

12, 13, 14

12, 13, 14

12, 13, 14

 

E-PROM Index control module

02, 03

 

≥ 02

02, 03

Table 5: Switch position for CO

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Measuring range:

 

 

0-150 ppm

 

 

 

 

S1/1

 

 

ON

 

 

 

 

S1/2

 

 

ON

 

 

 

 

S1/3

 

 

ON

 

 

 

 

S1/4

 

 

ON

 

 

 

 

S1/5

 

 

ON

 

 

 

 

S1/6

 

 

OFF

 

 

 

 

S1/7

 

 

OFF

 

 

 

 

S1/8

 

 

ON

 

 

 

 

E-PROM Index sensor head

 

12, 13, 14

 

 

 

 

E-PROM Index control module

 

≥ 02

 

 

Table 6: Switch position for HCl

Table 7: Switch position for ClO2

Measuring range:

0-100 ppm

S1/1

OFF

S1/2

ON

S1/3

ON

S1/4

ON

S1/5

ON

S1/6

ON

S1/7

ON

S1/8

ON

E-PROM Index sensor head

19

E-PROM Index control module

≥ 02

Measuring range:

0-0,5 ppm

S1/1

OFF

S1/2

ON

S1/3

ON

S1/4

OFF

S1/5

ON

S1/6

ON

S1/7

ON

S1/8

ON

E-PROM Index sensor head

17

E-PROM Index control module

≥ 02

17