Compur Statox 4120
Filter cartridge replacement
Remove tube
Remove filter cartridge and bearing
Clean air access opening
Replace filter cartridge and bearing
Replace tube
Generator cell replacement
Disconnect generator cell, remove tube to pump
Remove generator cell
Replace generator cell
Replace tube
Connect generator cell
6.2 Sensor head calibration
6.2.1 Using span gas
What you need: -Span gas
-Regulator
-Flow meter
-Gas calibration adapter
-Gas tube (clean, dry, chemical resistant)
-Diagnostic box
Caution:
The test gas adapter is not part of the ex-certified sensor head.
The gas flow should be adjusted to 500 ml/min (30 l/h).
Statox sensor heads working in the flow mode can be calibrated with the mounted flow adapter.
Keep the working mode gas flow.
Calibration adapter Flow adapter