I1 ATEX Intrinsic Safety

Certificate No.: BAS01ATEX1303XManual background II 1 G EEx ia IIC T4 (Ta = -60 °C to 70 °C)

-HART/Remote Display/Quick Connect/HART Diagnostics Manual background 1180

Table 2. Input Parameters

 

 

 

Loop / Power

Groups

Ui = 30V

All

Ii = 300 mA

All

Pi = 1.0W

All

Ci = 30nF

SuperModule/Quick Connect

Ci = 11.4nF

With PlantWeb or Junction Box Housing

Ci = 0

Remote Display

Li = 0

All Except Remote Display

Li = 60 µH

Remote Display

Special Conditions for Safe Use (X)

1.The apparatus, excluding the Types 3051 S-T and 3051 S-C (In-line and Coplanar SuperModule Platforms respectively), is not capable of withstanding the 500V test as defined in Clause 6.4.12 of EN 50020. This must be considered during installation.

2.The terminal pins of the Types 3051 S-T and 3051 S-C must be protected to IP20 minimum.

N1 ATEX Type n

Certificate No.: BAS01ATEX3304X Manual background II 3 G EEx nL IIC T4 (Ta = -40 °C TO 70 °C)