![Figure 4. Remote Mount Display wiring diagram](/images/new-backgrounds/90118/9011817x1.webp)
Quick Installation Guide
Rosemount 3051S
STEP 4 CONTINUED...
Intrinsic Safety Consideration: The transmitter assembly with remote display has been approved with Belden 3084A DeviceNet cable. Alternate cable may be used as long as the transmitter with remote display and cable is configured according to the installation control drawing or certificate. Refer to appropriate approval certificate or control drawing in Appendix B of the 3051S reference manual for remote cable IS requirements.
IMPORTANT
Do not apply power to the remote communications terminal. Follow wiring instructions carefully to prevent damage to system components.
Figure 4. Remote Mount Display wiring diagram
Junction Box Housing | Remote Mount Display |
|
|
(white) 24 AWG![Manual background](/images/new-backgrounds/90118/9011817xi3.webp)
(blue) 24 AWG
(black) 22 AWG
(red) 22 AWG
9