Quick Installation Guide
Rosemount 3095 MultiVariable
GATEX Type n
Certificate Number: BAS98ATEX3360X II 3 G
Ex nL IIC T5 (Tamb =
The apparatus is designed for connection to a remote temperature sensor such as a resistance temperature detection (RTD)
Special Conditions for Safe Use
The 3095, when fitted with the transient terminal block (order code B), are not capable of withstanding the 500 volts insulation test required by
HATEX Flameproof
Certificate Number: KEMA02ATEX2320X II 1/2 G Ex d IIC T5
T6
1180
Special Conditions for Safe Use (x):
The device contains a thin wall diaphragm. Installation, maintenance, and use shall take into account the environmental conditions to which the diaphragm will be subjected. The manufacturer’s instructions for installation and maintenance shall be followed in detail to assure safety during its expected lifetime.
PATEX Dust
Certificate Number: KEMA02ATEX2321 II 1 D Ex tD A20 T
V = 55 Vdc MAX I = 23 mA MAX IP66
1180
IECEx Certifications (HART)
4IECEx Intrinsic Safety
Certificate Number: IECEx BAS06.0070X Ex ia IIC T4
Ex ia IIC T5
Table 6. Input Parameters
HART I.S.
Ui = 30Vdc
Ii = 200 mAdc
Pi = 1.0 W
Ci = 12 nF
Li = 0
20