Rosemount 3095 Fieldbus

Quick Installation Guide

00825-0100-4716, Rev DA October 2009

Rosemount 3095 MultiVariable

Rosemount 3095 Fieldbus

Hazardous Locations Certifications

North American Certifications

FM Approvals

AExplosion Proof for Class I, Division 1, Groups B, C, and D. Dust-Ignition Proof for Class II/Class III, Division 1, Groups E, F, and G. Enclosure type NEMA 4X. Factory Sealed. Provides nonincendive RTD connections for Class I, Division 2, Groups A, B, C, and D.

JIntrinsically Safe for use in Class I, II and III, Division 1, Groups A, B, C, D, E, F, and G hazardous outdoor locations. Non-incendive for Class I, Division 2, Groups A, B, C, and D. Temperature Code T4. Factory Sealed.

For input parameters and installation see control drawing 03095-1020.

VFISCO for use in Class I, II and III, Division 1, Groups A, B, C, D, E, F, and G hazardous outdoor locations. Temperature Code T4. Factory Sealed.

For input parameters and installation see control drawing 03095-1020.

Canadian Standards Association (CSA)

CExplosion Proof for Class I, Division 1, Groups B, C, and D. Dust-Ignition Proof for Class II/Class III, Division 1, Groups E, F, and G. Factory Sealed. CSA enclosure Type 4X for indoor and outdoor hazardous locations. Suitable for Class I, Division 2, Groups A, B, C, and D. Install in accordance with Rosemount Drawing 03095-1024.

KIntrinsically Safe for Class I, Division 1, Groups A, B, C, and D. When installed in accordance with Rosemount Drawing 03095-1021. Temperature Code T3C.

WFISCO Field Device. Intrinsically Safe for Class I, Division 1, Groups A, B, C, and D. When installed in accordance with Rosemount Drawing 03095-1021. Temperature Code T3C.

European Certifications

F/T ATEX Intrinsic Safety Certificate Number: Baseefa 05ATEX0022X

North American Certifications II 1 G

EEx ia IIC T5 (Tamb = -45°C to 40°C) EEx ia IIC T4 (Tamb = -45°C to 70°C)

FM Approvals 1180

Table 8. Connection Parameters (Power/Signal Terminals)

Fieldbus (F Option)

FISCO (T Option)

Ui = 30V

Ui = 17.5V

Ii = 300 mA

Ii = 380 mA

Pi = 1.3 W

Pi = 5.32 W

Ci = 3.3 nF

Ci < 5 nF

Li = 0

Li = 10 µH

25

Page 25
Image 25
Emerson 00825-0100-4716 manual Rosemount 3095 Fieldbus Hazardous Locations Certifications, North American Certifications