Quick Installation Guide
Rosemount 3095 MultiVariable
PATEX Dust
Certificate Number: KEMA02ATEX2321
II 1 D T90°C
Imax = 23 mA IP66
1180
IECEx Certifications (Fieldbus)
4/Y IECEx Intrinsic Safety
Certificate Number: IECEx BAS05.0023X Ex ia IIC T4
Ex ia IIC T5
Table 11. Input Parameters |
| |
| Fieldbus I.S. | FISCO |
| Ui = 30Vdc | Ui = 17.5Vdc |
| Ii = 300 mAdc | Ii = 380 mAdc |
| Pi = 1.3 W | Pi = 5.32 W |
| Ci = 3.3 nF | Ci ≤ 5 nF |
| Li = 0 | Li ≤ 10 mH |
Table 12. RTD Terminals Entity Parameters
Uo = 30Vdc
Io = 19 mAdc
Po = 140 mW
The capacitance and either the Inductance or the Inductance to Resistance Ratio (L/R) of the load connected to the
Group | Capacitance | Inductance | or L/R Ratio |
| (F) | (mH) | H/Ohm |
IIC | 0.066 | 96 | 247 |
IIB | 0.56 | 365 | 633 |
IIA | 1.82 | 696 | 633 |
Conditions of Certification (X):
When fitted with the transient option, the apparatus is not capable of withstanding the 500V electrical strength test as defined in Clause 6.4.12 of IEC
5IECEx Type n
Certificate Number: IECEx BAS05.0024X Ex nC IIC T4
Ex nC IIC T5
Conditions of Certification (X):
When fitted with the transient option, the apparatus is not capable of withstanding the 500V electrical strength test as defined in Clause 8 of IEC
27