IECEx Certifications (Fieldbus)

Quick Installation Guide

00825-0100-4716, Rev DA October 2009

Rosemount 3095 MultiVariable

PATEX Dust

Certificate Number: KEMA02ATEX2321

Manual background II 1 D T90°C (-50°C Tamb 80°C) Vmax = 55 Vdc

Imax = 23 mA IP66

Manual background 1180

IECEx Certifications (Fieldbus)

4/Y IECEx Intrinsic Safety

Certificate Number: IECEx BAS05.0023X Ex ia IIC T4 (-45°C Ta 70°C)

Ex ia IIC T5 (-45°C Ta 40°C)

Table 11. Input Parameters

 

 

Fieldbus I.S.

FISCO

 

Ui = 30Vdc

Ui = 17.5Vdc

 

Ii = 300 mAdc

Ii = 380 mAdc

 

Pi = 1.3 W

Pi = 5.32 W

 

Ci = 3.3 nF

Ci 5 nF

 

Li = 0

Li 10 mH

Table 12. RTD Terminals Entity Parameters

Uo = 30Vdc

Io = 19 mAdc

Po = 140 mW

The capacitance and either the Inductance or the Inductance to Resistance Ratio (L/R) of the load connected to the 4-pin connector must not exceed the following values:

Group

Capacitance

Inductance

or L/R Ratio

 

(￿F)

(mH)

￿H/Ohm

IIC

0.066

96

247

IIB

0.56

365

633

IIA

1.82

696

633

Conditions of Certification (X):

When fitted with the transient option, the apparatus is not capable of withstanding the 500V electrical strength test as defined in Clause 6.4.12 of IEC 60079-11: 1999. This must be taken into account during installation.

5IECEx Type n

Certificate Number: IECEx BAS05.0024X Ex nC IIC T4 (-45°C Ta 70°C)

Ex nC IIC T5 (-45°C Ta 40°C) Ui = 55V dc max

Conditions of Certification (X):

When fitted with the transient option, the apparatus is not capable of withstanding the 500V electrical strength test as defined in Clause 8 of IEC 60079-15: 1987. This must be taken into account during installation.

27

Page 27
Image 27
Emerson 00825-0100-4716 manual IECEx Certifications Fieldbus, Rosemount 3095 MultiVariable, Quick Installation Guide