Parameter

Description

 

Min.

Max.

Default

No.

 

 

 

value

value

Ex-Factory

P130

 

Span 1 @ 1MHz

Measured attenuation

-

-

0.0

P131

 

Span 1 @ 3.3MHz

Measured attenuation

-

-

0.0

P132

 

Span 2 @ 1MHz

Measured attenuation

-

-

0.0

P133

 

Span 2 @ 3.3MHz

Measured attenuation

-

-

0.0

P134

 

Span 3 @ 1MHz

Measured attenuation

-

-

0.0

P135

 

Span 3 @ 3.3MHz

Measured attenuation

-

-

0.0

P140

 

Date for Span 1

 

-

-

0

P141

 

Date for Span 2

 

-

-

0

P142

 

Date for Span 3

 

-

-

0

P150

 

Lab Value 1

 

-

-

0

P151

 

Lab Value 2

 

-

-

0

P152

 

Lab Value 3

 

-

-

0

P160

 

Max % Solids

 

-

-

-

P200

 

PV Units

 

 

 

 

 

 

The PV units can be selected from a list

 

 

 

 

 

(%, kg/hr, lb/hr,g/l, ppm, None)

-

-

%

P201

 

Density Units

 

-

-

%

P240

 

Description

 

-

-

-

 

 

16 character user-defined description of application, say.

 

MSM400 MONIT

P241

 

Message

 

-

-

-

 

 

User set up message for HART info

 

32 Chars

 

P242

 

Tag Number - Control Unit

-

-

-

 

 

User-defined alpha-numeric entry up to 32 characters in

 

MSM40055

 

 

 

accordance with HART. May be entered by keypad.

 

 

 

P250

 

Start On

 

 

 

 

 

 

This parameter is used to set up the mode of start

 

 

 

 

 

for a desludge operation and is selected from a list

 

 

 

 

 

(Time, PV, Ext. Trig)

 

1

3

none

P251

 

Stop On

 

 

 

 

 

 

This parameter is used to set up the mode of stop

 

 

 

 

 

for a desludge operation and is selected from a list

 

 

 

 

 

(Time, PV, Ext. Trig)

 

1

3

none

P252

 

Do not Start if

 

 

 

 

 

 

This parameter is used to set up the mode 'do not start if'

 

 

 

 

 

for a desludge operation and is selected from a list

 

 

 

 

 

(Light sludge, Ext trigger)

1

3

not used

P253

 

Stop If select from list

 

 

 

 

 

This parameter is used to set up the mode 'stop if'

 

 

 

 

 

for a desludge operation and is selected from a list

 

 

 

 

 

(Light sludge, Ext trigger)

1

3

not used

P254

 

Start Time #1

 

 

 

 

 

 

The time of day that the first desludge cycle starts

00:00

23:59

7:00

P255

 

Interval #1

 

 

 

 

 

 

The time interval for the first desludge cycle

00:00

23:59

1:00

P256

 

Start Time #2

 

 

 

 

 

 

The time of day that the second desludge cycle starts

00:00

23:59

00:00

P257

 

Interval #2

 

 

 

 

 

 

The time interval for the first desludge cycle

00:00

23:59

00:00

P258

 

Max retries

 

 

Not implemented

 

P300

P

Sensor Serial Number

-

-

-

 

 

User-settable

 

 

 

 

P301

P

Sensor Type

 

-

-

-

P302

 

dB Damping

 

-

-

-

 

 

 

 

 

 

IP258

34

Page 34
Image 34
Emerson manual Parameter Description Min Max Default Value, MSM400 Monit