APPENDIX A2: FULL LIST OF FUNCTIONS
Parameter list and description
P=Parameter, | D=Diagnostic Display Parameter |
Parameter | Description |
No. |
|
P100 | Sensor Gap in mm. |
| This is a user entered parameter (optional). |
| This value is ignored when an alternative span |
| calibration is caried out (i.e. method two or three). |
P101 | Sludge type |
| The user can select from a list of sludge type to set |
| up the span for the calibration procedure (optional). |
| Each sludge type is associated with an attenuation |
| factor for each frequency. This value is ignored when |
| an alternative span calibration is caried out |
| (i.e. method two or three). |
P102 | dB Factor @ 1MHz |
| A user alterable parameter. This is the attenuation |
| number in dB/% at 1MHz. This is used for the span |
| calibration (method three). |
P103 | dB Factor @ 3.3MHz |
| A user alterable parameter. This is the attenuation |
| number in dB/% at 3.3MHz.This is used for the span |
| calibration (method three). |
P120 | Zero Ref @ 1MHz |
| This is the value in dBs (@ 1MHz) that is stored in |
| the control unit when a zero calibration is carried out. |
| This value is over written each time a zero calibration |
| is carried out. |
P121 | Zero Ref @ 3.3MHz |
| This is the value in dBs (@ 3.3MHz) that is stored in |
| the control unit when a zero calibration is carried out. |
| This value is over written each time a zero calibration |
| is carried out. |
P122 | Date of Zero Ref |
| This is automatically stored by the control unit when |
| the zero calibration is done. The actual date can be |
| |
P123 | Initial Zero Ref @ 1MHz dB’s |
| This is the value in dBs (@ 1MHz) that is stored in |
| the control unit when the initial zero calibration is |
| carried out.This value is compared with subsequent |
| zero calibrations.0 must be entered to reset or clear initial |
| zero value |
P124 | Initial Zero Ref @ 3.3MHz dB’s |
| This is the value in dBs (@ 3.3MHz) that is stored in |
| the control unit when the initial zero calibration is |
carried out.This value is compared with subsequent
zero calibrations.0 must be entered to reset or clear initial
Min | Max | Default |
value | value | |
0 | 1000 | 0.0 |
- | - | none |
0 | - | 0.000 |
0 | - | 0.000 |
- | - | 16 |
- | - | 16 |
- | - | |
- | - | 0 |
IP258