748213-S April 2002

1-8 SPECIFICATIONS1

a. Performance

 

 

Operating Range (Standard)

......... 0

to 5, 0 to 10, 0 to 25, 0 to 50, and 0 to 100% oxygen

Operating Range (Optional) ..........

0

to 1, 0 to 2.5, 0 to 5, 0 to 10, 0 to 25, 0 to 50, and 0 to 100%

 

oxygen

Response Time .............................

90% of fullscale, 20 seconds

Reproducibility...............................

0.01% oxygen or ±1% of fullscale, whichever is greater

Ambient Temperature Limits

......... 32°F (0°C) to 113°F (45°C)

Zero Drift........................................

±1% fullscale per 24 hours, provided that ambient temperature

 

does not change by more than 20°F (11.1°C)

 

±2.5% of fullscale per 24 hours with ambient temperature change

 

over entire range

Span Drift.......................................

±1% fullscale per 24 hours, provided that ambient temperature

 

does not change by more than 20°F (11.1°C)

 

±2.5% of fullscale per 24 hours with ambient temperature change

 

over entire range

b. Sample

 

Dryness .........................................

Sample dewpoint below 110°F (43°C), sample free of entrained

 

liquids.

Temperature Limits .......................

50°F (10°C) to 150°F (65°C)

Operating Pressure .......................

Maximum: 10 psig (68.9 kPa)

.......................................................

Minimum: 5 psig vacuum (34.5 kPa vacuum)

Flow Rate ......................................

50 cc/min. to 500 cc/min.

.......................................................

Recommended 250 ±20 cc/min.

Materials in Contact with Sample..

Glass, 316 stainless steel, titanium, Paliney No. 7, epoxy resin,

 

Viton-A, platinum, nickel, and MgF2

1Performance specifications are measured at recorder output and are based on constant sample pressure and deviation from set flow held to within 10% or 20 cc/min., whichever is smaller.

Rosemount Analytical Inc. A Division of Emerson Process Management

Description and Specifications

1-3