Model 755R
Instruction Manual
WARNING
TOXIC OR CORROSIVE HAZARD
The sample containment system must be carefully leak checked upon installation and before initial
Internal leaks resulting from failure to observe these precautions could result in personal injury or property damage.
For proper operation and safety, system leakage must be corrected, particularly before introduction of toxic or corrosive samples and/or application of electrical power.
To check system for leaks, liberally cover all fittings, seals, and other possible sources of leakage with suitable leak test liquid such as SNOOP (P/N 837801). Check for leak indicative bubbling or foaming. Leaks that are inaccessible to SNOOP application could evade detection by this method.
A BI CD E
(Rear terminal cover removed for clarity)
L1/HOT
L2/NEUT
GND
CUR VOLT
OUTPUT OUTPUT
+ - G + -
HG H
A.Sample outlet. 1/4” O.D. tube fitting.
B.Sample Inlet. 1/4” O.D. tube fitting.
C.5/8” diameter hole for optional Dual Alarm Cable. Cable supplied by customer, minimum 24 AWG.
D.5/8” diameter hole fitted with
E.13/16” diameter hole for Power Cable. Cable supplied by customer, 3 conductor, minimum 18 AWG.
F.TB1: Customer
G.TB2: Customer
H.Optional Dual Alarm connections.
I.Connections for Optional Remote Range Change.
Figure
Rosemount Analytical Inc. A Division of Emerson Process Management | Installation |