Instruction Manual

748213-S April 2002

2-4 LEAK TEST

WARNING

TOXIC OR CORROSIVE HAZARD

The sample containment system must be carefully leak checked upon installation and before initial start-up, during routine maintenance and any time the integrity of the sample containment system is broken, to ensure the system is in leak proof con- dition.

Internal leaks resulting from failure to observe these precautions could result in personal injury or property damage.

For proper operation and safety, system leakage must be corrected, particularly before introduction of toxic or corrosive samples and/or application of electrical power.

To check system for leaks, liberally cover all fittings, seals, and other possible sources of leakage with suitable leak test liquid such as SNOOP (P/N 837801). Check for leak indicative bubbling or foaming. Leaks that are inaccessible to SNOOP application could evade detection by this method.

A BI CD E

(Rear terminal cover removed for clarity)

L1/HOT

L2/NEUT

GND

CUR VOLT

OUTPUT OUTPUT

+ - G + -

HG H

A.Sample outlet. 1/4” O.D. tube fitting.

B.Sample Inlet. 1/4” O.D. tube fitting.

C.5/8” diameter hole for optional Dual Alarm Cable. Cable supplied by customer, minimum 24 AWG.

D.5/8” diameter hole fitted with liquid-tight gland for Recorder Output Cable. Cable supplied by customer, conductor, minimum 24 AWG.

E.13/16” diameter hole for Power Cable. Cable supplied by customer, 3 conductor, minimum 18 AWG.

F.TB1: Customer hook-up for Power.

G.TB2: Customer hook-up for Recorder Output.

H.Optional Dual Alarm connections.

I.Connections for Optional Remote Range Change.

Figure 2-2. Model 755R Rear Panel

Rosemount Analytical Inc. A Division of Emerson Process Management

Installation