Model 755R
Instruction Manual
c.Potentiometric Output
1.Insert RECORDER OUTPUT Selector Plug (See Figure
2.Connect leads of shielded recorder cable to “REC OUT +” and
3.Connect the output cable to the appro- priate terminals of the recorder or other potentiometric device:
a.For device with span of 0 to 10 mV, 0 to 100 mV, 0 to 1V, or 0 to 5V, connect cable directly to input terminals of the device, ensuring correct polarity and range selec- tion.
b.For a device with intermediate span (i.e., between the specified values), connect the cable to the device via a suitable external voltage divider (See Figure
d.Isolated Current Output (Optional)
1.Verify that the optional current output board appropriate to desired output is properly in place in its connector. See Figure
2.On I/O board, connect leads of shielded recorder cable to “CUR- RENT OUT+” and
3.Connect free end of output cable to input terminals of recorder or other
Current and voltage outputs may be utilized simultaneously if desired.
mA +
-
755R
+
-
Recorder
Analyzer
Note: Total series resistance of all devices is not to exceed 1000 ohms.
+Controller
-
+Remote - Indicator
Figure
Rosemount Analytical Inc. A Division of Emerson Process Management | Installation |