
Instruction Manual
Model 755R
RANGE % OXYGEN | RECOMMENDED ZERO | RECOMMENDED SPAN | |
STANDARD GAS | STANDARD GAS | ||
| |||
0 to 1 | Nitrogen | 0.9% O2, balance N2 | |
0 to 2.5 | Nitrogen | 2.3% O2, balance N2 | |
0 to 5 | Nitrogen | 4.5% O2, balance N2 | |
0 to 10 | Nitrogen | 9% O2, balance N2 | |
0 to 25 | Nitrogen | Air (20.93% O2) | |
0 to 50 | Nitrogen | 45% O2, balance N2 | |
0 to 100 | Nitrogen | 100% O2 |
Table
a.Oxygen Equivalent Value of Gases
For computation of background correc- tions, the analyzer response to each component of the sample must be shown. Table
The percentage oxygen equivalent of a gas is the instrument response to the given gas compared to the response to oxygen, assuming that both gases are supplied at the same pressure .
Example: Zero Based Range
In equation form: |
| |
%O2 Equivalent of Gas = |
| |
Analyzer Response to Gas | X 100% | |
Analyzer Response to O2 | ||
|
To select a random example from Table
The oxygen equivalent of a gas mixture is the sum of the contribution of the individual gas components.
At lower range limit, i.e., 0% oxygen, composition of sample is 80% CO2, 20% N2.
From Table
% oxygen equivalent of mixture =
0.8x
b.Computing Adjusted Settings for Zero and Span Controls
During instrument calibration, adjusted values may be required in setting the ZERO and SPAN controls to correct for the magnetic susceptibility of the background gas.
The quantities are defined as follows:
BGGst = Oxygen equivalent of background gas in standard gas (see Table
BGGs = Oxygen equivalent of background gas sample (see Table
OP = operating pressure. Unless special pressure corrections are to be made, the zero stan- dard, span standard and sample gases must all be admitted at the same pressure.
Rosemount Analytical Inc. A Division of Emerson Process Management |