Model 755R
Instruction Manual
a.Source Lamp
The simplest check of the source lamp is to verify that it is lit. Another check is done by removing the housing cover and viewing the lamp through the photocell alignment hole (see Figure
±0.2 VDC, the Power Supply Board must be replaced.
b.Photocell
If the photocells are on, observe through the photocell alignment hole. The image should be steady. Disconnect the line power and observe the image when you reconnect the power. It should come up from the side and seek a position that equally illuminates the photocells.
c.Suspension
Turn electrical power to instrument OFF. Remove optical bench assembly (see Figure
If the suspension has been changed, the cause may be improper operating condi- tions.
a.Source Lamp
Removal/Installation
The source lamp is held in the optical bench assembly by a set screw (see Fig- ure
The red line on the lamp base must align with the set screw (see Figure
Realign the photocell per Section
b.Photocell
Removal/Installation
Refer to Figure
Reverse the removal procedure for in- stallation. Align photocell (see below).
Alignment
The adjustments in this procedure are made on the Control Board. With zero gas flowing:
1.Place a digital voltmeter between the wiper of zero potentiometer (R13) and TP7 (ground). Adjust for 0 VDC.
2.Remove the voltmeter from R13 and place on R10 (see Figure
3.Remove the voltmeter from R10 and place on TP8. Move the photocell to obtain a DC voltage as close to 0 mV as possible, but no more than ±750 mV.
Rosemount Analytical Inc. A Division of Emerson Process Management | Maintenance and Service |