b.Pressure Requirements - General
Operating pressure limits are as follows: maximum, 10 psig (68.9 kPa); minimum, 5 psig vacuum (34.5 kPa vacuum).
CAUTION
RANGE LIMITATIONS
Operation outside the specified pressure limits may damage the detector, and will void the warranty.
The basic rule for pressure of sample and standard gases supplied to the inlet is to calibrate the analyzer at the same pressure that will be used during subsequent operation, and to maintain this pressure during operation. The arrangement required to obtain appropriate pressure control will depend on the application. When inputting sample or calibration gases, use the same pressure that will be used during subsequent operation. Refer to Section
c.Normal Operation at Positive Gauge Pressures
Normally, the sample is supplied to the analyzer inlet at a positive gauge pressure in the range of 0 to 10 psig (0 to 68.9 kPa).
CAUTION
HIGH PRESSURE GAS CYLINDERS
Pressure surges in excess of 10 psig during admission of sample or standard gases can damage the detector.
Maximum permissible operating pressure is 10 psig (68.9 kPa). To ensure against
The analyzer exhaust port is commonly vented directly to the atmosphere. Any change in barometric pressure results in a