Instruction Manual
Model 755R
equipment from the analyzer power cable | |
| (Refer to Figure |
WARNING
ELECTRICAL SHOCK HAZARD
For safety and proper performance, this instrument must be connected to a properly grounded
Cable connections for AC power, recorder output, and alarm output are shown in Installation Drawing, 654015, and are explained in the following sections.
a.Line Power Connection
The analyzer is supplied, as ordered, for operation on 115 VAC or 230 VAC, 50/60 Hz. Ensure that the power source con- forms to the requirements of the individual instrument, as noted on the
Electrical power is supplied to the ana- lyzer via a
If the analyzer is mounted in a protected rack or cabinet or on a bench, an accessory kit (P/N 654008) is available which provides a
b.Recorder Output Selection and Cable Connections
If a recorder, controller, or other output device is used, connect it to the analyzer via a number 22 or number 24 AWG
NOTE:
Route recorder cable through a separate cable gland (P/N 899329) or conduit not with power cable or alarm output cable.
Cable connections and output selection for potentiometric and
755R | Potentiometric |
Analyzer | Recorder |
| Input |
| Terminals |
Voltage Divider | (Verify polarity |
is correct) | |
(Customer Supplied) | |
Position of Recorder Output | Minimum Permissible |
Selector Plug | Resistance for R1 + R2 |
| (ohms) |
10 mV | 1K |
100 mV | 10K |
1 V | 100K |
5 V | 2K |
Figure
Installation | Rosemount Analytical Inc. A Division of Emerson Process Management |