Model 755R
Instruction Manual
The ALARM 1 and ALARM 2 setpoint adjustments (see Figure
Required dial settings for both setpoint adjustments may be determined from either Figure
•
•Required control setting = (desired alarm setpoint)(10)
fullscale span
Figure
Operating range, 0 to 5% oxygen
Desired ALARM 1 setpoint = 4% oxygen
Turn potentiometer R64 to 8
Desired ALARM 2 setpoint = 3% oxygen
Turn potentiometer R68 to 6
The desired deadband may be selected via the appropriate trimming potentiometer, R73, for ALARM 1 deadband adjustment and R78 for ALARM 2 deadband adjustment. For any setpoint, deadband is adjustable from 1% of fullscale (counterclockwise limit) to 20% of fullscale (clockwise limit). Deadband is essentially symmetrical with respect to setpoint.
The Current Output is set at the factor for 4 to 20 mA. If a 0 to 20 mA output is required, readjust both the zero and span potentiometers (R1 and R2) on the Current Output Board.
RANGE |
|
|
| PERCENTAGE OXYGEN READOUT |
|
| ||||||||||||||||||
% |
|
|
|
|
|
|
|
|
|
|
| versus |
|
|
|
|
|
|
|
| ||||
OXYGEN |
|
|
| ALARM SETPOINT DIAL READING |
|
| ||||||||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||||||||
|
|
|
|
|
|
|
| Percentage Oxygen Readout |
|
|
|
| ||||||||||||
| 0.0 | 0.1 | 0.2 | 0.3 | 0.4 | 0.5 | 0.6 | 0.7 | 0.8 | 0.9 | 1.0 | |||||||||||||
0 to 1 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||
0 | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 | ||||||||||||||
| ||||||||||||||||||||||||
|
|
|
|
|
|
|
|
| Setpoint Dial Reading |
|
|
|
|
|
| |||||||||
|
|
|
|
|
|
|
| Percentage Oxygen Readout |
|
|
|
| ||||||||||||
| 0.0 | 0.1 | 0.2 | 0.3 | 0.4 | 0.5 | 0.6 | 0.7 | 0.8 | 0.9 | 1.0 | |||||||||||||
0 to 2.5 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||
0 | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 | ||||||||||||||
| ||||||||||||||||||||||||
|
|
|
|
|
|
|
|
| Setpoint Dial Reading |
|
|
|
|
|
| |||||||||
|
|
|
|
|
|
|
| Percentage Oxygen Readout |
|
|
|
| ||||||||||||
0 to 5 | 0.0 | 0.5 | 1.0 | 1.5 | 2.0 | 2.5 | 3.0 | 3.5 | 4.0 | 4.5 | 5.0 | |||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||
| 0 | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 | |||||||||||||
|
|
|
|
|
|
|
|
| Setpoint Dial Reading |
|
|
|
|
|
| |||||||||
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||||||||
|
|
|
|
|
|
|
| Percentage Oxygen Readout |
|
|
|
| ||||||||||||
0 to 10 | 0 | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 | |||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||
| 0 | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 | |||||||||||||
|
|
|
|
|
|
|
|
| Setpoint Dial Reading |
|
|
|
|
|
| |||||||||
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||||||||
|
|
|
|
|
|
|
| Percentage Oxygen Readout |
|
|
|
| ||||||||||||
0 to 25 | 0 | 2.5 | 5 | 7.5 | 10 | 12.5 | 15 | 17.5 | 20 | 22.5 | 25 | |||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||
| 0 | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 | |||||||||||||
|
|
|
|
|
|
|
|
| Setpoint Dial Reading |
|
|
|
|
|
| |||||||||
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||||||||
|
|
|
|
|
|
|
| Percentage Oxygen Readout |
|
|
|
| ||||||||||||
| 0 | 5 | 10 | 15 | 20 | 25 | 30 | 35 | 40 | 45 | 50 | |||||||||||||
0 to 50 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||
0 | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 | ||||||||||||||
| ||||||||||||||||||||||||
|
|
|
|
|
|
|
|
| Setpoint Dial Reading |
|
|
|
|
|
| |||||||||
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||||||||
|
|
|
|
|
|
|
| Percentage Oxygen Readout |
|
|
|
| ||||||||||||
0 to 100 | 0 | 10 | 20 | 30 | 40 | 50 | 60 | 70 | 80 | 90 100 | ||||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||
| 0 | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 | |||||||||||||
|
|
|
|
|
|
|
|
| Setpoint Dial Reading |
|
|
|
|
|
| |||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Figure
Adjustments
Rosemount Analytical Inc. A Division of Emerson Process Management | Operation |