3-6 SELECTION OF SETPOINTS AND DEAD- BAND ON ALARM OPTION

The ALARM 1 and ALARM 2 setpoint adjustments (see Figure 3-1, page 3-3) are adjustable for any desired value from 1% to 100% of the fullscale analyzer span. The adjustment screws are graduated from 0 to 10.

Required dial settings for both setpoint adjustments may be determined from either Figure 3-2 or the appropriate equation that follows:

Zero-based operating range

Required control setting = (desired alarm setpoint)(10)

fullscale span

Figure 3-2 example:

Operating range, 0 to 5% oxygen

Desired ALARM 1 setpoint = 4% oxygen

Turn potentiometer R64 to 8

Desired ALARM 2 setpoint = 3% oxygen

Turn potentiometer R68 to 6

The desired deadband may be selected via the appropriate trimming potentiometer, R73, for ALARM 1 deadband adjustment and R78 for ALARM 2 deadband adjustment. For any setpoint, deadband is adjustable from 1% of fullscale (counterclockwise limit) to 20% of fullscale (clockwise limit). Deadband is essentially symmetrical with respect to setpoint.

3-7 CURRENT OUTPUT BOARD (OPTION)

The Current Output is set at the factor for 4 to 20 mA. If a 0 to 20 mA output is required, readjust both the zero and span potentiometers (R1 and R2) on the Current Output Board.

RANGE

 

 

 

PERCENTAGE OXYGEN READOUT

 

 

%

 

 

 

 

 

 

 

 

 

 

 

versus

 

 

 

 

 

 

 

 

OXYGEN

 

 

 

ALARM SETPOINT DIAL READING

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Percentage Oxygen Readout

 

 

 

 

 

0.0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1.0

0 to 1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0

1

2

3

4

5

6

7

8

9

10

 

 

 

 

 

 

 

 

 

 

Setpoint Dial Reading

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Percentage Oxygen Readout

 

 

 

 

 

0.0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1.0

0 to 2.5

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0

1

2

3

4

5

6

7

8

9

10

 

 

 

 

 

 

 

 

 

 

Setpoint Dial Reading

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Percentage Oxygen Readout

 

 

 

 

0 to 5

0.0

0.5

1.0

1.5

2.0

2.5

3.0

3.5

4.0

4.5

5.0

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0

1

2

3

4

5

6

7

8

9

10

 

 

 

 

 

 

 

 

 

Setpoint Dial Reading

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Percentage Oxygen Readout

 

 

 

 

0 to 10

0

1

2

3

4

5

6

7

8

9

10

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0

1

2

3

4

5

6

7

8

9

10

 

 

 

 

 

 

 

 

 

Setpoint Dial Reading

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Percentage Oxygen Readout

 

 

 

 

0 to 25

0

2.5

5

7.5

10

12.5

15

17.5

20

22.5

25

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0

1

2

3

4

5

6

7

8

9

10

 

 

 

 

 

 

 

 

 

Setpoint Dial Reading

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Percentage Oxygen Readout

 

 

 

 

 

0

5

10

15

20

25

30

35

40

45

50

0 to 50

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0

1

2

3

4

5

6

7

8

9

10

 

 

 

 

 

 

 

 

 

 

Setpoint Dial Reading

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Percentage Oxygen Readout

 

 

 

 

0 to 100

0

10

20

30

40

50

60

70

80

90 100

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0

1

2

3

4

5

6

7

8

9

10

 

 

 

 

 

 

 

 

 

Setpoint Dial Reading

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Figure 3-2. Dial Settings for Alarm Setpoint

Adjustments