Instruction Manual
Model 755R
e.Output Connections and Initial Setup for Dual Alarm Option
If so ordered, the analyzer is factory equipped with alarm output. Alternatively, the alarm feature is obtainable by subsequent installation of the 654019 Alarm Kit.
The alarm output provides two sets of relay contacts for actuation of alarm and/or
REQUIREMENT | TYPICAL CONNECTIONS |
| ||
| No. 1 | NO |
| N |
|
| COM |
| H |
|
| NC | Alarm Bell | 115 VAC |
| RESET | or Lamp |
| |
Low Alarm, |
|
| ||
|
|
|
| |
| NO |
|
| |
|
| COM |
|
|
|
| NC |
|
|
| RESET |
|
| |
| No. 2 |
|
|
|
| No. 1 | NO |
|
|
|
| COM |
|
|
|
| NC |
|
|
High Alarm, | RESET |
|
| |
| NO |
| N | |
|
|
| ||
|
| COM |
| |
|
|
| H | |
|
|
|
| |
|
| NC | Alarm Bell | 115 VAC |
| RESET | or Lamp |
| |
|
|
| ||
| No. 2 |
|
|
|
| No. 1 | NO |
|
|
|
| COM |
|
|
|
| NC |
|
|
Low Control | RESET | Solenoid |
| |
Limit, |
|
|
| |
| NO | Valve |
| |
|
| |||
| COM |
| H | |
|
|
| ||
|
| NC |
| N |
| RESET |
| 115 VAC | |
|
|
| ||
| No. 2 |
|
|
|
REQUIREMENT | TYPICAL CONNECTIONS |
| ||
|
|
| Solenoid |
|
| No. 1 | NO | Valve |
|
|
| COM |
| H |
|
| NC |
| N |
| RESET |
| 115 VAC | |
Low Control |
|
| ||
Limit, |
| NO |
|
|
|
|
| ||
| COM |
|
| |
|
|
|
| |
|
| NC |
|
|
| RESET |
|
| |
| No. 2 |
|
|
|
Lower | No. 1 | NO |
| N |
| COM |
| ||
Low Alarm |
|
| H | |
|
|
| ||
Indicator, |
| NC | Alarm Bell | 115 VAC |
RESET | or Lamp |
| ||
| Solenoid |
| ||
|
|
|
| |
|
| NO | Valve |
|
Low Control, |
| COM |
| H |
| NC |
| N | |
RESET |
| 115 VAC | ||
|
|
| ||
| No. 2 |
|
|
|
|
|
| Solenoid |
|
| No. 1 | NO | Valve |
|
High Control, |
| COM |
| H |
| NC |
| ||
|
| N | ||
|
|
| ||
| RESET |
| 115 VAC | |
|
|
| ||
Higher |
| NO |
| N |
| COM |
| H | |
High Alarm |
|
| ||
| NC | Alarm Bell | 115 VAC | |
Indicator, |
| |||
|
| or Lamp |
| |
RESET |
| |||
|
| |||
| No. 2 |
|
|
|
Figure
Note the following recommendations:
•A fuse should be inserted into the line between the
•If the alarm contacts are connected to any device that produces radio fre-
quency interference (RFI), it should be arc suppressed. The 858728 Arc Suppressor is recommended.
•If at all possible, the analyzer should operate on a different AC power source, to avoid RFI.