Instruction Manual

748213-S April 2002

Model 755R

e.Output Connections and Initial Setup for Dual Alarm Option

If so ordered, the analyzer is factory equipped with alarm output. Alternatively, the alarm feature is obtainable by subsequent installation of the 654019 Alarm Kit.

The alarm output provides two sets of relay contacts for actuation of alarm and/or process-control functions. Leads from the customer-supplied external alarm system connect to terminals on the 654019 Alarm Assembly (See Figure 2-5 below and Interconnect Drawing 654014).

REQUIREMENT

TYPICAL CONNECTIONS

 

 

No. 1

NO

 

N

 

 

COM

 

H

 

 

NC

Alarm Bell

115 VAC

 

RESET

or Lamp

 

Low Alarm,

 

 

 

 

 

 

Fail-Safe

 

NO

 

 

 

 

COM

 

 

 

 

NC

 

 

 

RESET

 

 

 

No. 2

 

 

 

 

No. 1

NO

 

 

 

 

COM

 

 

 

 

NC

 

 

High Alarm,

RESET

 

 

Fail-Safe

 

NO

 

N

 

 

 

 

 

COM

 

 

 

 

H

 

 

 

 

 

 

NC

Alarm Bell

115 VAC

 

RESET

or Lamp

 

 

 

 

 

No. 2

 

 

 

 

No. 1

NO

 

 

 

 

COM

 

 

 

 

NC

 

 

Low Control

RESET

Solenoid

 

Limit,

 

 

 

 

NO

Valve

 

Fail-Safe

 

 

 

COM

 

H

 

 

 

 

 

NC

 

N

 

RESET

 

115 VAC

 

 

 

 

No. 2

 

 

 

REQUIREMENT

TYPICAL CONNECTIONS

 

 

 

 

Solenoid

 

 

No. 1

NO

Valve

 

 

 

COM

 

H

 

 

NC

 

N

 

RESET

 

115 VAC

Low Control

 

 

Limit,

 

NO

 

 

Fail-Safe

 

 

 

 

COM

 

 

 

 

 

 

 

 

NC

 

 

 

RESET

 

 

 

No. 2

 

 

 

Lower

No. 1

NO

 

N

 

COM

 

Low Alarm

 

 

H

 

 

 

Indicator,

 

NC

Alarm Bell

115 VAC

Fail-Safe

RESET

or Lamp

 

 

Solenoid

 

 

 

 

 

 

 

NO

Valve

 

Low Control,

 

COM

 

H

 

NC

 

N

Fail-Safe

RESET

 

115 VAC

 

 

 

 

No. 2

 

 

 

 

 

 

Solenoid

 

 

No. 1

NO

Valve

 

High Control,

 

COM

 

H

Fail-Safe

 

NC

 

 

 

N

 

 

 

 

RESET

 

115 VAC

 

 

 

Higher

 

NO

 

N

 

COM

 

H

High Alarm

 

 

 

NC

Alarm Bell

115 VAC

Indicator,

 

 

 

or Lamp

 

Fail-Safe

RESET

 

 

 

 

No. 2

 

 

 

Figure 2-5. Relay Terminal Connections for Typical Fail-Safe Applications

Note the following recommendations:

A fuse should be inserted into the line between the customer-supplied power supply and the alarm relay terminals on the Alarm Relay Assembly.

If the alarm contacts are connected to any device that produces radio fre-

quency interference (RFI), it should be arc suppressed. The 858728 Arc Suppressor is recommended.

If at all possible, the analyzer should operate on a different AC power source, to avoid RFI.

2-8