SECTION 4 THEORY
Oxygen is strongly paramagnetic while most other common gases are weakly diamagnetic. The paramagnetism of oxygen may be regarded as the capability of an oxygen molecule to become a temporary magnet when placed in a magnetic field. This is analogous to the magnetization of a piece of soft iron. Diamagnetic gases are analogous to
With the Model 755R, the volume magnetic susceptibility of the flowing gas sample is sensed in the detector/magnet assembly. As shown in the functional diagram of Figure
Because of the “magnetic buoyancy” effect, the spheres of the test body are subjected to displacement forces, resulting in a displacement torque that is proportional to the volume magnetic susceptibility of the gas surrounding the test body.
Measurement is accomplished by a
In effect, each sphere is wound with a
The restoring current is automatically maintained at the correct level by an
The output current from the dual photocell is equal to the difference between the signals developed by the two halves of the photocell. This difference, which constitutes the error signal, is applied to the input of an amplifier circuit that provides the restoring current.
When the test body is in null position, both halves of the photocell are equally illuminated, the error signal is zero, and the amplifier is unequal. This condition results in application of an error signal to the input of the amplifier circuit. The resultant amplifier output signal is routed through the current loop, thus creating the electromagnetic forces required to restore the test body to null position.
Additionally, the output from the amplifier is conditioned as required to drive the digital display, and recorder if used. The electronic circuitry involved is described briefly in Section