The power supply circuitry on the Remote Range Board 646004 must be jumpered for the correct line voltage, either 115 VAC or 230 VAC. See Drawings # 656081 (Table 2) and 646090 for correct jumper locations.
On the Remote Range Board, an additional option exists: for using either the
To use an external supply:
1.Remove the E to F jumper (DWG 646090).
2.Apply the external 12 V to
3.Program the remote controller to pull the range bits,
To use the internal 12 V supply:
1.Verify the E to F jumper is in place.
2.Connect the controller's common to
NOTE
DO NOT connect anything to
3.Connect
Remember that you are dealing with inverse logic and not normal binary addresses. Also, this process switches the recorder output only, and does not affect the front panel dis- play.
|
|
| Hex | ||||
Range 1 |
| 1 | 1 | 1 | 0 |
| E |
|
|
|
|
|
|
|
|
Range 2 |
| 1 | 1 | 0 | 1 |
| D |
|
|
|
|
|
|
|
|
Range 3 |
| 1 | 0 | 1 | 1 |
| B |
|
|
|
|
|
|
|
|
Range 4 |
| 0 | 1 | 1 | 1 |
| 7 |
|
|
|
|
|
|
| |
Note: 1 = 12 V, 0 << 1 V. |
|
|
|
|
| ||
Table | Remote Range Switching Truth Table |
|
Installation | Rosemount Analytical Inc. A Division of Emerson Process Management |