Fisher 244ER, 244EH manual Configuring Several Transmitters Identically

Models: 244EH 244ER

1 64
Download 64 pages 54.53 Kb
Page 30
Image 30
Configuring Several Transmitters Identically

Rosemount Models 244EH and 244ER PC-Programmable Temperature Transmitters

Configuring Several Transmitters Identically

NOTE

Each configuration parameter is explained in detail in the on-line help provided with the software. To access help, click the HELP button on the main configuration window (see Figure 3-1).

5.Click the SEND TO XMTR button to upload configuration information to the transmitter. The software will indicate whether the configuration is successful or unsuccessful. If the configuration succeeds, then go to step Step 6. If the configuration does not succeed, then refer to Section 4: Maintenance and Troubleshooting for a list of possible causes and repeat Step 5 until the transmitter configuration is successful.

6.Click the SAVE TO FILE button to save the configuration information for use with other transmitters.

NOTE

To ease the process of later finding a specific configuration file, name saved files according to a logical system and preserve the default .244 filename extension.

7.Click the CANCEL button to close the main configuration window.

To configure several Model 244E transmitters identically, perform the following procedure:

1.Configure a single transmitter according to “Configuring a Single Transmitter” on page 3-3and save the configuration to a file.

2.Select OPEN from the FILE menu to open the saved configuration file.

3.Connect a new transmitter to the Configuration Interface.

4.Click the SEND TO XMTR button to upload the configuration information to the transmitter. The software will report whether the configuration is successful or unsuccessful. If the configuration succeeds, then go to Step 5. If the configuration does not succeed, then refer to Section 4: Maintenance and Troubleshooting for a list of possible causes and repeat Step 4 until the transmitter configuration is successful.

5.Repeat Steps 3 and 4 until all transmitters are configured.

6.Click the CANCEL button to close the main configuration screen.